Difference between revisions of "Stepper Recipes"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
(→‎Stepper 3 (ASML DUV): link to edge bead removal)
 
(20 intermediate revisions by 2 users not shown)
Line 32: Line 32:
 
! width="300" |Comments
 
! width="300" |Comments
 
|-
 
|-
|SPR955CM0.9
+
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM0.9]]
 
|3 krpm/30”
 
|3 krpm/30”
 
|95°C/60”
 
|95°C/60”
Line 46: Line 46:
   
 
|-
 
|-
  +
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-0.9]]
|SPR955CM0.9
 
 
|3 krpm/30”
 
|3 krpm/30”
 
|95°C/60”
 
|95°C/60”
Line 61: Line 61:
   
 
|-
 
|-
  +
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-0.9]]<br>
|SPR955CM0.9<br>
 
  +
[[Media:CEM365iS-Contrast-Enhancement-Datasheet.pdf|CEM365iS]]
CEM365iS
 
 
|3 krpm/30”<br>
 
|3 krpm/30”<br>
 
5 krpm/30”
 
5 krpm/30”
Line 90: Line 90:
   
 
|-
 
|-
|SPR955CM-1.8
+
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-1.8]]
 
|4 krpm/30”
 
|4 krpm/30”
 
|90°C/90”
 
|90°C/90”
Line 104: Line 104:
   
 
|-
 
|-
|SPR955CM-1.8
+
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-1.8]]
 
|4 krpm/30”
 
|4 krpm/30”
 
|90°C/90”
 
|90°C/90”
Line 118: Line 118:
   
 
|-
 
|-
  +
|[[Media:SPR220-Positive-Resist-Datasheet.pdf|SPR220-3.0]]
|SPR220-3.0
 
 
|2.5 krpm/30”
 
|2.5 krpm/30”
 
|115°C/90”
 
|115°C/90”
Line 132: Line 132:
   
 
|-
 
|-
  +
|[[Media:SPR220-Positive-Resist-Datasheet.pdf|SPR220-7.0]]
|SPR220-7.0
 
 
|3.5 krpm/45”
 
|3.5 krpm/45”
 
|115°C/120”
 
|115°C/120”
Line 168: Line 168:
 
! width="300" |Comments
 
! width="300" |Comments
 
|-
 
|-
  +
|[[Media:AZ5214-Negative-Resist-Datasheet.pdf|AZ5214]]
|AZ5214
 
 
|6 krpm/30”
 
|6 krpm/30”
 
|95°C/60”
 
|95°C/60”
Line 182: Line 182:
   
 
|-
 
|-
  +
|[[Media:AZnLOF5510-Negative-Resist-Datasheet.pdf|nLOF5510]]
|nLOF5510
 
 
|3 krpm/30”
 
|3 krpm/30”
 
|90°C/60”
 
|90°C/60”
Line 198: Line 198:
   
 
|-
 
|-
  +
|[[Media:AZnLOF2020-Negative-Resist-Datasheet.pdf|nLOF2020]]
|nLOF2020
 
 
|4 krpm/30”
 
|4 krpm/30”
 
|110°C/60”
 
|110°C/60”
Line 235: Line 235:
 
! width="300" |Comments
 
! width="300" |Comments
 
|-
 
|-
|SPR955CM-0.9
+
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-0.9]]
 
|3 krpm/30”
 
|3 krpm/30”
 
|95°C/90”
 
|95°C/90”
Line 249: Line 249:
   
 
|-
 
|-
|SPR955CM-0.9
+
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-0.9]]
 
|3 krpm/30”
 
|3 krpm/30”
 
|95°C/90”
 
|95°C/90”
Line 262: Line 262:
   
 
|-
 
|-
|SPR955CM-1.8
+
|[[Media:SPR955-Positive-Resist-Datasheet.pdf|SPR955CM-1.8]]
 
|4 krpm/30”
 
|4 krpm/30”
 
|95°C/90”
 
|95°C/90”
Line 286: Line 286:
 
| align="left" |
 
| align="left" |
 
|-
 
|-
  +
|[[Media:SPR220-Positive-Resist-Datasheet.pdf|SPR220-3.0]]
|SPR220-3.0
 
 
|2.5 krpm/30”
 
|2.5 krpm/30”
 
|115°C/90”
 
|115°C/90”
Line 299: Line 299:
   
 
|-
 
|-
  +
|[[Media:SPR220-Positive-Resist-Datasheet.pdf|SPR220-7.0]]
|SPR220-7.0
 
 
|3.5 krpm/45”
 
|3.5 krpm/45”
 
|115°C/120”
 
|115°C/120”
Line 315: Line 315:
   
 
|-
 
|-
  +
|[[Media:AXP4000pb-Datasheet.pdf|AZ4210]]
|AZ4210
 
 
|step1:500rpm/5", step2:4krpm/45”
 
|step1:500rpm/5", step2:4krpm/45”
 
|95°C/60"
 
|95°C/60"
Line 321: Line 321:
 
|'''0.75"'''
 
|'''0.75"'''
 
|0
 
|0
|
+
|
   
|AZ400K:DI=1:4
+
|AZ400K:DI=1:4
 
|60"
 
|60"
 
| align="left" |
 
| align="left" |
Line 349: Line 349:
 
! width="300" |Comments
 
! width="300" |Comments
 
|-
 
|-
  +
|[[Media:AZnLOF5510-Negative-Resist-Datasheet.pdf|nLOF5510]]
|nLOF5510
 
 
|3 krpm/30”
 
|3 krpm/30”
 
|90°C/60”
 
|90°C/60”
Line 365: Line 365:
   
 
|-
 
|-
  +
|[[Media:AZ5214-Negative-Resist-Datasheet.pdf|AZ5214]]
|AZ5214
 
 
|6 krpm/30”
 
|6 krpm/30”
 
|95°C/60”
 
|95°C/60”
Line 379: Line 379:
   
 
|-
 
|-
  +
|[[Media:AZnLOF2020-Negative-Resist-Datasheet.pdf|nLOF2020]]
|nLOF2020
 
 
|4 krpm/30”
 
|4 krpm/30”
 
|110°C/60”
 
|110°C/60”
Line 394: Line 394:
 
|-
 
|-
   
|NR9-1000PY
+
|[[Media:NR9-1000PY-revA.pdf|NR9-1000PY]]
 
|3 krpm/30”
 
|3 krpm/30”
 
|135°C/180” lid down
 
|135°C/180” lid down
Line 411: Line 411:
   
 
=[[Stepper 3 (ASML DUV)]]=
 
=[[Stepper 3 (ASML DUV)]]=
  +
{| class="wikitable"
  +
!Stepper 3: Table of Contents
  +
|-
  +
|
  +
#[[Stepper_Recipes#Process_Control_Data|Process Control Data]]
  +
#[[Stepper Recipes#Photomasks .26 Job Programming|Photomasks & Job Programming]]
  +
#[[Stepper Recipes#Anti-Reflective Coatings|Anti-Reflective Coatings]]
  +
#[[Stepper Recipes#Positive Resist .28ASML DUV.29|Positive Resist]]
  +
#[[Stepper Recipes#Negative Resist .28ASML DUV.29|Negative Resist]]
  +
#[[Stepper Recipes#Other Lithography Processes .28ASML DUV.29|Other Lithography Processes]]
  +
##Edge Bead Removal
  +
##Lift-off Processes
  +
|}
  +
  +
==Process Control Data==
  +
  +
*''The Process Group regularly measures data on lithography Critical Dimension ("CD") and Wafer-stage Particulate Contamination for this tool, using a sensitive lithography process that will reveal small changes in Dose repeatability and wafer flatness.''
  +
*[https://docs.google.com/spreadsheets/d/1xW1TFH_QjPMWl9T1jiKzwmYe4B2wg7KY-nqOKUoXttI/edit#gid=1804752281 '''Plots of CD Repeatability''']
  +
*[https://docs.google.com/spreadsheets/d/1xW1TFH_QjPMWl9T1jiKzwmYe4B2wg7KY-nqOKUoXttI/edit#gid=0 '''Data for CD Uniformity and Particulate Contamination''']
  +
  +
==Photomasks & Job Programming==
  +
You can find info on making photomasks, CAD files for alignment marks, and other Job programming tools on the main Tool page, here:
  +
  +
[[Stepper 3 (ASML DUV)#Design%20Tools|Stepper 3 (ASML DUV) > Design Tools]]
   
 
==Anti-Reflective Coatings==
 
==Anti-Reflective Coatings==
Bottom Anti-reflective coatings (aka. BARC or AR Coating) are, in general, used for the ASML stepper. LOL2000 and PMGI can also be used as under layers.
+
Bottom Anti-reflective coatings (aka. BARC or AR Coating) are, in general, used for the ASML stepper. LOL2000 and PMGI can also be used as under layers. BARC layers will increase the process tolerance, especially for features ≤ wavelength (248nm).
  +
  +
For example, without BARC you might find that 200nm features only resolve with focus of -0.2µm to -0.3µm, while with BARC the process tolerance may increase to -0.1µm to -0.4µm, which also increases tolerance to wafer flatness/PR spin uniformity.
   
==='''[https://wiki.nanotech.ucsb.edu/wiki/images/0/07/DUV42P-Anti-Reflective-Coating.pdf DUV-42P]'''===
+
====[https://wiki.nanotech.ucsb.edu/wiki/images/0/07/DUV42P-Anti-Reflective-Coating.pdf <big>DUV-42P-6</big>]====
 
''(replacement for AR2)''
 
''(replacement for AR2)''
   
*spin coat at 3500rpm for a 670Å thick coating.
+
*spin coat at 2500rpm for optimal anti-reflective properties (~60nm).
  +
**Mistake: this was previously written as 3500rpm.
 
*Bake at 220°C for 60s on a hotplate.
 
*Bake at 220°C for 60s on a hotplate.
 
*This AR coating is removed via oxygen plasma.
 
*This AR coating is removed via oxygen plasma.
*This ARC can be etched on [https://wiki.nanotech.ucsb.edu/wiki/index.php/ICP_Etching_Recipes#Photoresist_and_ARC_Etching ICP#1], [https://wiki.nanotech.ucsb.edu/wiki/index.php/ICP_Etching_Recipes#Photoresist_and_ARC_etching ICP#2] or [https://wiki.nanotech.ucsb.edu/wiki/index.php/RIE_Etching_Recipes#Photoresist_and_ARC RIE#5]
+
*This ARC can be etched on [https://wiki.nanotech.ucsb.edu/w/index.php?title=ICP_Etching_Recipes#Photoresist_and_ARC_Etching_.28Panasonic_1.29 ICP#1], [https://wiki.nanotech.ucsb.edu/w/index.php?title=ICP_Etching_Recipes#Photoresist_and_ARC_etching_.28Panasonic_2.29 ICP#2], [https://wiki.nanotech.ucsb.edu/w/index.php?title=RIE_Etching_Recipes#Photoresist_and_ARC_.28RIE_5.29 RIE#5], [https://wiki.nanotech.ucsb.edu/wiki/ICP_Etching_Recipes#Photoresist_.26_ARC_.28Fluorine_ICP_Etcher.29 FL-ICP] or even [[Ashers (Technics PEII)|Technics PEii ashers]].
*Datasheet: '''[https://wiki.nanotech.ucsb.edu/wiki/images/0/07/DUV42P-Anti-Reflective-Coating.pdf DUV-42P]'''
+
*Datasheet: '''[https://wiki.nanotech.ucsb.edu/wiki/images/0/07/DUV42P-Anti-Reflective-Coating.pdf DUV-42P-6]'''
   
==='''[https://wiki.nanotech.ucsb.edu/wiki/images/a/af/DS-K101-304-Anti-Reflective-Coating.pdf DS-K101-304]'''===
+
====[https://wiki.nanotech.ucsb.edu/wiki/images/a/af/DS-K101-304-Anti-Reflective-Coating.pdf <big>DS-K101-304</big>]====
   
*spin at 5000rpm and bake at 185C for 60s.
+
*Spin at 1500rpm and bake at 185°C for 60sec
  +
**Approx 40nm for best anti-reflective properties
**~20nm thickness
 
  +
**Mistake: previously written as 5000rpm, ~20nm thickness (from DS-K101-307)
 
*This AR coating develops away and undercuts in AZ300MIF.
 
*This AR coating develops away and undercuts in AZ300MIF.
 
**For isolated lines, this can cause them to lift-off by undercutting the resist.
 
**For isolated lines, this can cause them to lift-off by undercutting the resist.
Line 451: Line 479:
 
! width="300" |Comments
 
! width="300" |Comments
 
|-
 
|-
|UV6-0.7 <br> (replaced by UV6-0.8)
+
|UV6-0.7 <br> (replaced by [[Media:UV6-Positive-Resist-Datasheet.pdf|UV6-0.8]])
   
 
|3.5 krpm/30”
 
|3.5 krpm/30”
Line 468: Line 496:
   
 
|-
 
|-
  +
|[[Media:UV210-Positive-Resist-Datasheet.pdf|UV210-0.3]]
|UV210-0.3
 
 
|5.0 krpm/30”
 
|5.0 krpm/30”
 
|135°C/60”
 
|135°C/60”
Line 483: Line 511:
   
 
|-
 
|-
  +
|[[Media:UV210-Positive-Resist-Datasheet.pdf|UV210-0.3]]
|UV210-0.3
 
 
|3.0 krpm/30”
 
|3.0 krpm/30”
 
|135°C/90”
 
|135°C/90”
Line 496: Line 524:
 
*'''Annular Illumination'''
 
*'''Annular Illumination'''
 
*NA 0.63, Sigma_o 0.8, Sigma_i 0.5
 
*NA 0.63, Sigma_o 0.8, Sigma_i 0.5
  +
|-
 
  +
|[[Media:UV26-Positive-Resist-Datasheet.pdf|UV26-2.5]]
  +
|''Available but no recipes characterized''
  +
|135°C/90s
  +
|~2.5 µm
  +
|approx. 40
  +
|approx. +0.8
  +
|110°C/90s
  +
|AZ300MiF
  +
|Unknown
  +
|Users must run your own development/FEM's.
 
|}
 
|}
   
Line 516: Line 554:
 
! width="300" |Comments
 
! width="300" |Comments
 
|-
 
|-
  +
|[[Media:UVN-30 - Negative-Resist-Datasheet - Apr 2004.pdf|UVN30-0.8]]
|UVN30-0.8
 
 
|3.5 krpm/30”
 
|3.5 krpm/30”
 
|110°C/60”
 
|110°C/60”
Line 525: Line 563:
 
|Not Used
 
|Not Used
 
|AZ300MIF
 
|AZ300MIF
  +
|Approx. 20sec
|55-60 sec
 
  +
(not thoroughly calibrated)
 
| align="left" |
 
| align="left" |
*Replaced UVN2300, not identical
+
*Replaced UVN2300, not identical. Dev time was 55-60 sec.
   
 
|}
 
|}
   
== Other Lithography Porcess (ASML DUV) ==
+
==Other Lithography Processes (ASML DUV)==
  +
* [[ASML DUV: Edge Bead Removal via Photolithography|Photolithographic Edge Bead Removal]]
+
*[[ASML DUV: Edge Bead Removal via Photolithography|DUV Photolithographic Edge Bead Removal]]
* DUV Lift-Off Process
 
  +
*[[Lift-Off with DUV Imaging + PMGI Underlayer|DUV Lift-Off Process with PMGI Underlayer]]

Latest revision as of 17:03, 21 March 2024

Back to Lithography Recipes.

Below is a listing of stepper lithography recipes. Stepper 1 and Stepper 2 are i-line systems with good piece handling capabilities.  Stepper 3 is a DUV (248nm) system primarily used for full 100mm wafers.  DUV resists do not work for i-line and i-line resists do not work for DUV.  Based on your sample reflectivity, absorption (or whether or not you use an ARC layer), and surface topography, the exposure time / focus offset parameters may vary.

This listing is a guideline to get you started. For critical lithography steps, you should run your own exposure and/or focus array to determine the proper parameters.

The recipes are tabulated to give you the values of the key parameters you will need to establish your recipe. Underlayers such as LOL2000 or PMGI can be used on the stepper systems.  See the underlayer datasheets for details.   Post develop bakes (not listed) are used to make the resist more etch resistant and depend on subsequent processes. Care should be taken with post development bakes as resist reflow can occur. Unless otherwise noted, all exposures are done on flat, silicon wafers.

Parameters are indicated in separate tables for each stepper system.

Stepper 1 (GCA 6300)

Multiply the GCA 6300 exposure times by 0.30 to get a starting exposure time for the GCA Autostep200 system.

Positive Resist (GCA 6300)

Unless otherwise noted, bakes are on hot plates. For recipes with CEM, the CEM is spun on after the first resist bake, exposure is then done, and the CEM is rinsed off with DI water before the PEB. CEM generally improves resolution and process tolerance at the expense of higher exposure time.

Resist Spin Cond. Bake Thickness Exposure Time Focus Offset PEB Developer Developer Time Comments
SPR955CM0.9 3 krpm/30” 95°C/60” ~ 0.9 um 1.2” 0 110°C/60” AZ300MIF 60"
SPR955CM-0.9 3 krpm/30” 95°C/60” ~ 0.9 um 3.0” 4 110°C/60” AZ300MIF 60"
SPR955CM-0.9

CEM365iS

3 krpm/30”

5 krpm/30”

95°C/90” ~ 0.9 um 2.2” -10 110°C/60” AZ300MIF 60"
SPR950-0.8 4 krpm/30” 95°C/60” ~ 0.8 um 1.0” 0 105°C/60” AZ300MIF 60"
SPR955CM-1.8 4 krpm/30” 90°C/90” ~ 1.8 um 2.3” 0 110°C/90” AZ300MIF 60"
SPR955CM-1.8 4 krpm/30” 90°C/90” ~ 1.8 um 1.7” -5 110°C/90” AZ300MIF 60"
SPR220-3.0 2.5 krpm/30” 115°C/90” ~ 2.7 um 2.4” 10 115°C/90” AZ300MIF 60"
SPR220-7.0 3.5 krpm/45” 115°C/120” ~ 7.0 um 4.5” 0 *50°C/60”

115°C/90”

AZ300MIF 120"
  • 1.0 um isolated lines; 1.25 um isolated spaces
    • Let sample sit in air for 20 minutes before PEB, step to 50°C for 60” first, then 115°C
  • See SPR220-7 Data File

Negative Resist (GCA 6300)

Unless otherwise noted, bakes are on hot plate. All flood exposures are done in broadband light using any contact aligner. Also, because the tone is negative, a shorter first exposure time will result in more undercut, which is desirable for single-layer lift-off processes. Under these conditions more develop time will also give more undercut.

Resist Spin Cond. Bake Thickness Exposure Time Focus Offset PEB Flood Developer Developer Time Comments
AZ5214 6 krpm/30” 95°C/60” ~ 1.0 um 0.2” 0 110°C/60” 60" AZ300MIF 60"
  • 0.7 um res. possible 
nLOF5510 3 krpm/30” 90°C/60” ~ 0.93 um 0.74” -6 110°C/60” 0 AZ300MIF 60"
  • 0.5 um line openings good dense or isolated
  • Use heated 1165 stripper for removal or lift-off
  • See nLOF5510 data file
nLOF2020 4 krpm/30” 110°C/60” ~ 2 um 0.55” -6 110°C/60” 0 AZ300MIF 90"
  • ~ .85 um line opening/lift-off good. Isolated mesas can be smaller.
  • Use heated 1165 stripper for removal or lift-off Sensetive to PEB temp.
  • See nLOF2020 Data File

Stepper 2 (AutoStep 200)

Positive Resist (AutoStep 200)

Unless otherwise noted, bakes are on hot plates. For recipes with CEM, the CEM is spun on after the first resist bake, exposure is then done, and the CEM is rinsed off with DI water before the PEB. CEM generally improves resolution and process tolerance at the expense of higher exposure time.

NOTE: The bolded exposure times were found by multiplying the exposure times from the GCA 6300 system by 0.30. They should be sued as a starting point. You will need to do an exposure array to get precise times for the Autostep system. In general, the resolution achievable is ~ 100 nm smaller for the Autostep200 system.

Resist Spin Cond. Bake Thickness Exposure Time Focus Offset PEB Developer Developer Time Comments
SPR955CM-0.9 3 krpm/30” 95°C/90” ~ 0.9 um 0.35” 0 110°C/90” AZ300MIF 60”
SPR955CM-0.9 3 krpm/30” 95°C/90” ~ 0.9 um 0.8” 0 110°C/90” AZ300MIF 60”
  • 0.5um holes
SPR955CM-1.8 4 krpm/30” 95°C/90” ~ 1.8 um 0.4” -1 110°C/90” AZ300MIF 60”
SPR950-0.8 4 krpm/30” 95°C/60” ~ 0.8 um 0.30” 0 105°C/60” AZ300MIF 60"
SPR220-3.0 2.5 krpm/30” 115°C/90” ~ 2.7 um 0.72” 10 115°C/90” AZ300MIF 60"
  • 0.5 um isolated lines
SPR220-7.0 3.5 krpm/45” 115°C/120” ~ 7.0 um 1.35" 0 *50°C/60”

115°C/90”

AZ300MIF 120"
  • 1.0 um isolated lines; 1.25 um isolated spaces
    • Let sample sit in air for 20 minutes before PEB, step to 50°C for 60” first, then 115°C
AZ4210 step1:500rpm/5", step2:4krpm/45” 95°C/60" ~ 2.1 um 0.75" 0 AZ400K:DI=1:4 60"
  • 2.0 um dense holes

Negative Resist (AutoStep 200)

Unless otherwise noted, bakes are on hot plate. All flood exposures are done in broadband light using any contact aligner. Also, because the tone is negative, a shorter first exposure time will result in more undercut, which is desirable for single-layer lift-off processes. Under these conditions more develop time will also give more undercut.

NOTE: The bolded exposure times were found by multiplying the exposure times from the GCA 6300 system by 0.30. They should be sued as a starting point. You will need to do an exposure array to get precise times for the Autostep system. In general, the resolution achievable is ~ 100 nm smaller for the Autostep200 system.

Resist Spin Cond. Bake Thickness Exposure Time Focus Offset PEB Flood Developer Developer Time Comments
nLOF5510 3 krpm/30” 90°C/60” ~ 0.93 um .25” -1 110°C/60” 0 AZ300MIF 60”
AZ5214 6 krpm/30” 95°C/60” ~ 1.0 um 0.06” 0 110°C/60” 60" AZ300MIF 60"
  • 0.7 um res. possible
nLOF2020 4 krpm/30” 110°C/60” ~ 2 um 0.17” -6 110°C/60” 0 AZ300MIF 90"
  • ~ .85 um line opening/lift-off good. Isolated mesas can be smaller.
  • Use heated 1165 stripper for removal or lift-off Sensetive to PEB temp.
NR9-1000PY 3 krpm/30” 135°C/180” lid down ~ 1.2 um 0.92” 0 115°C/120” lid down 0 AZ300MIF 20"

Stepper 3 (ASML DUV)

Stepper 3: Table of Contents
  1. Process Control Data
  2. Photomasks & Job Programming
  3. Anti-Reflective Coatings
  4. Positive Resist
  5. Negative Resist
  6. Other Lithography Processes
    1. Edge Bead Removal
    2. Lift-off Processes

Process Control Data

Photomasks & Job Programming

You can find info on making photomasks, CAD files for alignment marks, and other Job programming tools on the main Tool page, here:

Stepper 3 (ASML DUV) > Design Tools

Anti-Reflective Coatings

Bottom Anti-reflective coatings (aka. BARC or AR Coating) are, in general, used for the ASML stepper. LOL2000 and PMGI can also be used as under layers. BARC layers will increase the process tolerance, especially for features ≤ wavelength (248nm).

For example, without BARC you might find that 200nm features only resolve with focus of -0.2µm to -0.3µm, while with BARC the process tolerance may increase to -0.1µm to -0.4µm, which also increases tolerance to wafer flatness/PR spin uniformity.

DUV-42P-6

(replacement for AR2)

  • spin coat at 2500rpm for optimal anti-reflective properties (~60nm).
    • Mistake: this was previously written as 3500rpm.
  • Bake at 220°C for 60s on a hotplate.
  • This AR coating is removed via oxygen plasma.
  • This ARC can be etched on ICP#1, ICP#2, RIE#5, FL-ICP or even Technics PEii ashers.
  • Datasheet: DUV-42P-6

DS-K101-304

  • Spin at 1500rpm and bake at 185°C for 60sec
    • Approx 40nm for best anti-reflective properties
    • Mistake: previously written as 5000rpm, ~20nm thickness (from DS-K101-307)
  • This AR coating develops away and undercuts in AZ300MIF.
    • For isolated lines, this can cause them to lift-off by undercutting the resist.
  • Increase bake temperature to reduce undercut rate.
  • Datasheet: DS-K101-304

Positive Resist (ASML DUV)

Please see section above for anti-reflection coatings, which are usually used with the DUV Stepper.

Resist Spin Cond. Bake Thickness Exposure Dose(mj) Focus Offset PEB Developer Developer Time Comments
UV6-0.7
(replaced by UV6-0.8)
3.5 krpm/30” 135°C/60” 630nm 17 -0.2 135°C/90” AZ300MIF 45”
  • 200nm dense line/space
  • NA 0.57, Sigma 0.75
  • Eo ~ 5.5mj
  • UV6-0.7 was discontinued, we now stock UV6-0.8
UV210-0.3 5.0 krpm/30” 135°C/60” 230nm 20 -0.1 135°C/90” AZ300MIF 45"
  • 150nm dense line/space
  • Annular Illumination
  • NA 0.63, Sigma_o 0.8, Sigma_i 0.5
UV210-0.3 3.0 krpm/30” 135°C/90” 260nm 85 -0.2 135°C/90” AZ300MIF 80”
  • 170nm isolated holes
  • Annular Illumination
  • NA 0.63, Sigma_o 0.8, Sigma_i 0.5
UV26-2.5 Available but no recipes characterized 135°C/90s ~2.5 µm approx. 40 approx. +0.8 110°C/90s AZ300MiF Unknown Users must run your own development/FEM's.

Negative Resist (ASML DUV)

Please see section above for anti-reflection coatings, which are usually used with the DUV Stepper.

Resist Spin Cond. Bake Thickness Exposure Dose (mj) Focus Offset PEB Flood Developer Developer Time Comments
UVN30-0.8 3.5 krpm/30” 110°C/60” ~550nm 27 +0.15 105°C/60” Not Used AZ300MIF Approx. 20sec

(not thoroughly calibrated)

  • Replaced UVN2300, not identical. Dev time was 55-60 sec.

Other Lithography Processes (ASML DUV)