Sputtering Recipes
Revision as of 14:34, 15 April 2014 by Biljana (talk | contribs) (→Ta{{sub|2}}O{{sub|5}} deposition (IBD))
Back to Vacuum Deposition Recipes.
Sputter 1 (Custom)
Sputter 2 (SFI Endeavor)
Al Deposition (Sputter 2)
AlNx Deposition (Sputter 2)
Au Deposition (Sputter 2)
TiO2 Deposition (Sputter 2)
Sputter 3 (AJA ATC 2000-F)
The recipes below are given as starting points from data obtained in the nanofab. For critical depositions, calibrations are recommended.
Material | P(mT) | Pow(W) | Sub(W) | T(C) | Ar | N2 | O2 | Height-Tilt | Rate(nm/min) | Stress(MPa) | Rs(uOhm-cm) | n@633nm | k@633nm | Data Below | Comment |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX | XXX |
Ni and Ta Deposition (Sputter 3)
Sputter 4 (AJA ATC 2200-V)
The recipes below are given as starting points from data obtained in the nanofab. For critical depositions, calibrations are recommended.
Material | P(mT) | Pow(W) | Sub(W) | T(C) | Ar | N2 | O2 | Height-Tilt | Rate(nm/min) | Stress(MPa) | Rs(uOhm-cm) | n@633nm | k@633nm | Data Below | Comment |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
W | 3 | 300 | 0 | 50 | 45 | 0 | 0 | H2.75-T5 | 11.5 | -150 to 150 | 11 | - | - | Yes | Jeremy Watcher |
TiW | 4.5 | 300 | 0 | 75 | 45 | 0 | 0 | H2.75-T5 | 9.5 | -150 to 150 | 60 | - | - | Yes | 10%Ti by Wt |
W-TiW Deposition (Sputter 4)
Sputter 5 (Lesker AXXIS)
Ion Beam Deposition (Veeco NEXUS)
IBD Calibrations Spreadsheet - Records of historical film depositions (rates, indices), Uniformity etc.
All users are required to enter their calibration deps (simple test deps only)
SiO2 deposition (IBD)
- SiO2 Standard Recipe
- SiO2 Data 2014 (Deposition Rate, Refractive Index, Stress, HF etch rate )
- SiO2 Data April 2014
- SiO2 Thickness uniformity 2014
- Deposition Rate: ≈ 5.5 nm/min (users must calibrate this prior to critical deps)
- Refractive Index: ≈ 1.487
- Stress ≈ -373MPa (compressive)
Si3N4 deposition (IBD)
- Si3N4 Standard Recipe
- Si3N4 Data 2014 (Deposition Rate, Refractive Index, Stress, HF etch rate )
- Si3N4 Data April 2014
- Si3N4 Thickness uniformity 2014
- Deposition Rate ≈ 4.1nm/min (users must calibrate this prior to critical deps)
- Refractive Index ≈ 1.972
- Stress ≈ -1583MPa (compressive)
Ta2O5 deposition (IBD)
- Ta2O5 Standard Recipe
- Ta2O5 Data 2014 (Deposition Rate, Refractive Index, Stress, HF etch rate )
- Ta2O5 Data March 2014
- Ta2O5 Thickness uniformity 2014
- Deposition Rate ≈ 8.5nm/min (users must calibrate this prior to critical deps)
- Refractive Index ≈ 2.174
- Stress ≈ -228MPa (compressive)
TiO2 deposition (IBD)
- TiO2 Standard Recipe
- TiO2 Data 2014 (Deposition Rate, Refractive Index, Stress, HF etch rate )
- TiO2 Data March 2014
- TiO2 Thickness uniformity 2014
- Deposition Rate ≈ 1.4nm/min (users must calibrate this prior to critical deps)
- Refractive Index ≈ 2.346
- Stress ≈ -442MPa (compressive)