Difference between revisions of "Sputtering Recipes"
Jump to navigation
Jump to search
Line 14: | Line 14: | ||
=[[Sputter 4 (AJA ATC 2200-V)]]= |
=[[Sputter 4 (AJA ATC 2200-V)]]= |
||
== W Deposition (Sputter 4) == |
== W Deposition (Sputter 4) == |
||
− | *[[media:|W Deposition Recipe]] |
+ | *[[media:23-Tungsten_Sputtering_Film-Sputter-4.pdf|W Deposition Recipe]] |
=[[Sputter 5 (Lesker AXXIS)]]= |
=[[Sputter 5 (Lesker AXXIS)]]= |
Revision as of 16:10, 4 November 2013
Back to Vacuum Deposition Recipes.
Sputter 1 (Custom)
Sputter 2 (SFI Endeavor)
Al Deposition (Sputter 2)
AlNx Deposition (Sputter 2)
Au Deposition (Sputter 2)
TiO2 Deposition (Sputter 2)
Sputter 3 (AJA ATC 2000-F)
Sputter 4 (AJA ATC 2200-V)
W Deposition (Sputter 4)
Sputter 5 (Lesker AXXIS)
Ion Beam Deposition (Veeco NEXUS)
IBD Calibrations Spreadsheet - Records of historical film depositions (rates, indices), Uniformity etc.
All users are required to enter their calibration deps (simple test deps only)
SiO2 deposition (IBD)
- Refractive Index: ≈1.485
- Rate: ≈6.1nm/min (users must calibrate this prior to critical deps)
- Stress ≈ -320MPa (compressive)
Si3N4 deposition (IBD)
- Refractive Index ≈ 2.01
- Rate ≈ 4.5nm/min (users must calibrate this prior to critical deps)
- Stress ≈ -1756MPa (compressive)
Ta2O5 deposition (IBD)
- Refractive Index ≈ 2.10
- Rate ≈7.8nm/min (users must calibrate this prior to critical deps)
- Stress ≈ -140MPa (compressive)
TiO2 deposition (IBD)
- Refractive Index ≈ 2.3-2.4
- Rate ≈ 1.8nm/min
- Recipe: TiO2_dep