Difference between revisions of "RIE 2 (MRC)"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
(Created page with "{{tool|{{PAGENAME}} |picture=RIE2.jpg |type = Dry Etch |super= Don Freeborn |phone= 805-893-3918x216 |location=Bay 3 |email=freeborn@ece.ucsb.edu |description = RIE #2 Methane / …")
 
(SOP rev)
 
(17 intermediate revisions by 5 users not shown)
Line 2: Line 2:
 
|picture=RIE2.jpg
 
|picture=RIE2.jpg
 
|type = Dry Etch
 
|type = Dry Etch
|super= Don Freeborn
+
|super= Lee Sawyer
|phone= 805-893-3918x216
+
|phone= 805-893-2123
|location=Bay 3
+
|location=Bay 2
|email=freeborn@ece.ucsb.edu
+
|email=lee_sawyer@ucsb.edu
|description = RIE #2 Methane / Hydrogen-Based System
+
|description = RIE #2 Methane/Hydrogen-Based System
 
|manufacturer = Materials Research Corporation (MRC)
 
|manufacturer = Materials Research Corporation (MRC)
 
|materials =  
 
|materials =  
 +
|toolid=25
 
}}  
 
}}  
 +
==About==
  
= About  =
 
  
 
This is a Materials Research Corporation RIE-51 parallel plate, 13.56 Mhz system used primarily for the etching of InP with CH<sub>4</sub>/H<sub>2</sub>/Ar gases, although it can be used to etch As- and Sb-based III-V compounds and a variety of II-VI semiconductors as well. For Al-containing compounds and II-VI compounds, high bias power is required. Tool features include: six inch diameter water cooled cathode/substrate platform, pyrex cylinder for plasma confinement and gas flow control, adjustable cathode-anode spacing, fixed bias or power control and HeNe laser etch monitor with chart recorder. It is diffusion pumped and has no loadlock. Various etching applications have included: in-plane lasers/facets, InP-based HBTs, FET gate recessing, InP-based quantum microcavities, Bragg-Fresnel x-ray lenses and waveguides.  
 
This is a Materials Research Corporation RIE-51 parallel plate, 13.56 Mhz system used primarily for the etching of InP with CH<sub>4</sub>/H<sub>2</sub>/Ar gases, although it can be used to etch As- and Sb-based III-V compounds and a variety of II-VI semiconductors as well. For Al-containing compounds and II-VI compounds, high bias power is required. Tool features include: six inch diameter water cooled cathode/substrate platform, pyrex cylinder for plasma confinement and gas flow control, adjustable cathode-anode spacing, fixed bias or power control and HeNe laser etch monitor with chart recorder. It is diffusion pumped and has no loadlock. Various etching applications have included: in-plane lasers/facets, InP-based HBTs, FET gate recessing, InP-based quantum microcavities, Bragg-Fresnel x-ray lenses and waveguides.  
Line 17: Line 18:
 
RIE of InP and related compounds can be achieved with a hydride-based process chemistry of methane/hydrogen with an etching mechanism due to a "reverse" metalorganic CVD reaction. Because both etching and deposition occur simultaneously, it is important to use the proper gas flows and to periodically remove any polymer reaction by-products deposited on the non-etched (mask) surfaces. (This system has an additional flow circuit in order to bleed in small amounts, &lt;1 sccm, of O<sub>2</sub>). Alternatively, one can perform cyclic etching between MHA and O<sub>2</sub> to keep polymer formation to a minimum. With this technique selectivity is quite high and anisotropic etching can be achieved. While a metal, dielectric or photoresist may be used as a mask, photoresist should only be used at low bias voltages in order to avoid mask pattern distortions due to reflow. A precoat etch should be done before etching to condition the chamber.  
 
RIE of InP and related compounds can be achieved with a hydride-based process chemistry of methane/hydrogen with an etching mechanism due to a "reverse" metalorganic CVD reaction. Because both etching and deposition occur simultaneously, it is important to use the proper gas flows and to periodically remove any polymer reaction by-products deposited on the non-etched (mask) surfaces. (This system has an additional flow circuit in order to bleed in small amounts, &lt;1 sccm, of O<sub>2</sub>). Alternatively, one can perform cyclic etching between MHA and O<sub>2</sub> to keep polymer formation to a minimum. With this technique selectivity is quite high and anisotropic etching can be achieved. While a metal, dielectric or photoresist may be used as a mask, photoresist should only be used at low bias voltages in order to avoid mask pattern distortions due to reflow. A precoat etch should be done before etching to condition the chamber.  
  
= Detailed Specifications =
+
==Detailed Specifications==
  
*Etch gases include: CH<sub>4</sub>, H<sub>2</sub>, Ar and O<sub>2</sub>  
+
*Etch gases include: CH<sub>4</sub>, H<sub>2</sub>, Ar and O<sub>2</sub>
*Low 1 E -6 ultimate chamber pressure  
+
*Low 1 E -6 ultimate chamber pressure
*13.56 Mhz excitation frequency  
+
*13.56 Mhz excitation frequency
*Sample size limited to approximately 2 inches  
+
*Sample size limited to approximately 2 inches
*HeNe and IR laser monitoring for endpoint  
+
*HeNe and IR laser monitoring for endpoint
*Automatic tuning network  
+
*Automatic tuning network
*DC Bias or RF power control  
+
*DC Bias or RF power control
*Masking materials include: Ni, SiON, photoresist (limited to low bias/power)  
+
*Masking materials include: Ni, SiON, photoresist (limited to low bias/power)
 
*Typical etch conditions for InGaAsP:  
 
*Typical etch conditions for InGaAsP:  
**75 mT (CH<sub>4</sub>/H<sub>2</sub>/Ar&nbsp;: 4/20/10 sccm)  
+
**75 mT (CH<sub>4</sub>/H<sub>2</sub>/Ar&nbsp;: 4/20/10 sccm)
**450v bias  
+
**450v bias
 
**~ 45 nm/min. etch rate
 
**~ 45 nm/min. etch rate
 +
 +
==Documentation==
 +
 +
* [https://wiki.nanotech.ucsb.edu/w/images/8/82/RIE_2_SOP_Rev_B.pdf RIE #2 Standard Operating Procedure]

Latest revision as of 09:46, 13 September 2020

RIE 2 (MRC)
RIE2.jpg
Tool Type Dry Etch
Location Bay 2
Supervisor Lee Sawyer
Supervisor Phone (805) 893-2123
Supervisor E-Mail lee_sawyer@ucsb.edu
Description RIE #2 Methane/Hydrogen-Based System
Manufacturer Materials Research Corporation (MRC)
Dry Etch Recipes
Sign up for this tool


About

This is a Materials Research Corporation RIE-51 parallel plate, 13.56 Mhz system used primarily for the etching of InP with CH4/H2/Ar gases, although it can be used to etch As- and Sb-based III-V compounds and a variety of II-VI semiconductors as well. For Al-containing compounds and II-VI compounds, high bias power is required. Tool features include: six inch diameter water cooled cathode/substrate platform, pyrex cylinder for plasma confinement and gas flow control, adjustable cathode-anode spacing, fixed bias or power control and HeNe laser etch monitor with chart recorder. It is diffusion pumped and has no loadlock. Various etching applications have included: in-plane lasers/facets, InP-based HBTs, FET gate recessing, InP-based quantum microcavities, Bragg-Fresnel x-ray lenses and waveguides.

RIE of InP and related compounds can be achieved with a hydride-based process chemistry of methane/hydrogen with an etching mechanism due to a "reverse" metalorganic CVD reaction. Because both etching and deposition occur simultaneously, it is important to use the proper gas flows and to periodically remove any polymer reaction by-products deposited on the non-etched (mask) surfaces. (This system has an additional flow circuit in order to bleed in small amounts, <1 sccm, of O2). Alternatively, one can perform cyclic etching between MHA and O2 to keep polymer formation to a minimum. With this technique selectivity is quite high and anisotropic etching can be achieved. While a metal, dielectric or photoresist may be used as a mask, photoresist should only be used at low bias voltages in order to avoid mask pattern distortions due to reflow. A precoat etch should be done before etching to condition the chamber.

Detailed Specifications

  • Etch gases include: CH4, H2, Ar and O2
  • Low 1 E -6 ultimate chamber pressure
  • 13.56 Mhz excitation frequency
  • Sample size limited to approximately 2 inches
  • HeNe and IR laser monitoring for endpoint
  • Automatic tuning network
  • DC Bias or RF power control
  • Masking materials include: Ni, SiON, photoresist (limited to low bias/power)
  • Typical etch conditions for InGaAsP:
    • 75 mT (CH4/H2/Ar : 4/20/10 sccm)
    • 450v bias
    • ~ 45 nm/min. etch rate

Documentation