MVD - Wafer Coating - Process Traveler
Jump to navigation
Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Non-stick FDTS coating on the MVD
Used to make the wafer surface hydrophobic, used in nanoimprinting for coating the imprint master.
Use vapor coating in a clean-dry environment process as follows:
- Use Technics PEII etcher for 2 minutes with O2 at 0.300T/100W first to “activate” the silicon surface for reaction with the FDTS.
- Use Standard FDTS program on the MVD tool for this.
- Hot plate bake at ~100C for 2 minutes following deposition.
- H2O Contact Angle ~ 110 Degrees (measured with Goniometer)
Removal: of FDTS Coating
Technics PEii O2 etching removes this coating effectively, eg. 2min at the standard 0.300T/100W.