Difference between revisions of "DUV Flood Expose"
Jump to navigation
Jump to search
(System model and manufacturer info) |
|||
Line 1: | Line 1: | ||
− | <nowiki> |
+ | <nowiki><nowiki>Insert non-formatted text here</nowiki><nowiki><nowiki>Insert non-formatted text here</nowiki></nowiki></nowiki>{{tool|{{PAGENAME}} |
|picture=DUV.jpg |
|picture=DUV.jpg |
||
|type = Lithography |
|type = Lithography |
||
Line 6: | Line 6: | ||
|location=Bay 6 |
|location=Bay 6 |
||
|email=lee_sawyer@ucsb.edu |
|email=lee_sawyer@ucsb.edu |
||
− | |description = Deep UV Flood |
+ | |description = 1000W Deep UV Flood Exposure System (LS-150X-10C2) and Illumination Controller (2130-C2) |
− | |manufacturer = AB Manufacturing |
+ | |manufacturer = Bachur & Associates/AB Manufacturing |
|materials = |
|materials = |
||
}} |
}} |
Revision as of 16:35, 2 April 2019
<nowiki>Insert non-formatted text here<nowiki>Insert non-formatted text here</nowiki></nowiki>
|
About
This unit consists of a collimated deep ultraviolet (DUV) light source (mercury gas discharge short arc lamp) and power supply. The substrate is placed on a rotating chuck (not presently working) and is exposed by opening a timer-controlled shutter.
Materials that are exposed are primarily spun-on thin films such as PMMA, PMGI, etc.
Detailed Specifications
- DUV wavelengths are 200-260 nm; the lamp power is limited to 1000 watts and can operate in either constant intensity or constant power mode
- The full spectrum of Hg emission wavelengths from 200-450 nm is present on the sample
- Exposure can be performed on a 4" wafer
- Lamps are nominally rated for 400 hours
- A reset timer was added in Feb. 1995 to limit "on" time (ie. provide for auto shut-off) for bulb life conservation