Difference between revisions of "Wet Benches"

From UCSB Nanofab Wiki
Jump to: navigation, search
(Created page with "In the lab we have a total of 16 wet benches. They are divided into the following 8 types: *4 Solvent Cleaning Benches *3 Develop Benches *3 Spin Coat Benches *2 [[T…")
 
(Develop Benches)
Line 14: Line 14:
   
 
=Develop Benches=
 
=Develop Benches=
  +
{{tool|{{PAGENAME}}
  +
|picture=DevelopBench.jpg
  +
|type = Wet Processing
  +
|super= Aidan Hopkins
  +
|phone=(805)839-3918x208
  +
|location=?
  +
|email=hopkins@ece.ucsb.edu
  +
|description = Custom Developer Bench
  +
|manufacturer = Pure Aire Corporation
  +
|materials =
  +
}}
  +
== About ==
  +
  +
The facility contains 3 polypropylene Developing Benches for photolithographic development using TMAH / TEAH / KOH based developers. All is actively neutralized. The benches consist of two sinks with DI water hook ups and industrial water plenum flushes. Laminar flow fume hoods with hepa filtration. Nitrogen guns are also on both sides of the bench for drying samples. Digital hotplates are included for resist post exposure bakes or post develop hard bakes. POLOS spray puddle develop etch systems are included in the benches. All processes are done in user-supplied glassware. Two people may use the bench at a time and protective gear such as proper gloves must be used. Developers supplied by the laboratory include: AZ300MIF (0.26N TMAH developer), MF701 (0.24N TMAH developer), AZ400K (KOH based developer), pre-diluted AZ400K:DI 1:4. Other developers should be OK'd by laboratory staff prior to use in the laboratory.
  +
  +
== Deatailed Specifications ==
  +
  +
*2 Sinks, DI water
  +
*2 Nitrogen guns for sample drying
  +
*'''NO SOLVENTS or STRIPPERS''' in bench
  +
*Class 100
  +
*Hot plates for post bakes
  +
*POLOS Spray-Puddle-Develop-Rinse-Dry systems
   
 
=Spin Coat Benches=
 
=Spin Coat Benches=

Revision as of 07:38, 12 July 2012

In the lab we have a total of 16 wet benches. They are divided into the following 8 types:


Solvent Cleaning Benches

Develop Benches

Wet Benches
DevelopBench.jpg
Tool Type Wet Processing
Location ?
Supervisor Aidan Hopkins
Supervisor Phone (805) 893-2343
Supervisor E-Mail hopkins@ece.ucsb.edu
Description Custom Developer Bench
Manufacturer Pure Aire Corporation


About

The facility contains 3 polypropylene Developing Benches for photolithographic development using TMAH / TEAH / KOH based developers. All is actively neutralized. The benches consist of two sinks with DI water hook ups and industrial water plenum flushes. Laminar flow fume hoods with hepa filtration. Nitrogen guns are also on both sides of the bench for drying samples. Digital hotplates are included for resist post exposure bakes or post develop hard bakes. POLOS spray puddle develop etch systems are included in the benches. All processes are done in user-supplied glassware. Two people may use the bench at a time and protective gear such as proper gloves must be used. Developers supplied by the laboratory include: AZ300MIF (0.26N TMAH developer), MF701 (0.24N TMAH developer), AZ400K (KOH based developer), pre-diluted AZ400K:DI 1:4. Other developers should be OK'd by laboratory staff prior to use in the laboratory.

Deatailed Specifications

  • 2 Sinks, DI water
  • 2 Nitrogen guns for sample drying
  • NO SOLVENTS or STRIPPERS in bench
  • Class 100
  • Hot plates for post bakes
  • POLOS Spray-Puddle-Develop-Rinse-Dry systems

Spin Coat Benches

Toxic Corrosive Benches