Unaxis VLR Etch - Process Control Data

From UCSB Nanofab Wiki
Revision as of 11:20, 21 April 2022 by Pakala (talk | contribs) (edited recipe on Unaxis Process control data)
Jump to navigation Jump to search
UnderConstruction.jpg

Work In Progress

This article is still under construction. It may contain factual errors. Content is subject to change.


Data - InP Ridge Etch (Unaxis VLR)

PECVD SiO2 hardmask, patterned on Stepper #2 (AutoStep 200) & Panasonic ICP #1

InP Ridge Etch: °C, mT, 800W/125W, Cl2=6.3, H2=12.7, Ar=2sccm, time=1min30sec (90sec)

Sample Size: 1x1cm, ~30-40% SiO2 masking (NingC's pattern). Silicon carrier, no adhesive.

Date Sample# Etch Rate (nm/min) Etch Selectivity (InP/SiO2) Comments SEM Images
3/30/22 NP_Unaxis_02 1413 14.6 [1] [2]
3/9/22 NP_Unaxis_01 1297 15.3 [1] [2]