==Detailed Specifications==
 
==Detailed Specifications==
  
*ICP Power: ??? W
 
*RF Power: ??? W
 
 
*Temperature Range: –40°C to +400°C
 
*Temperature Range: –40°C to +400°C
 
*Gases Available: CH<sub>4</sub>, H<sub>2</sub>, Ar, Cl<sub>2</sub>, BCl<sub>3</sub>, SF<sub>6</sub>, SiCl<sub>4</sub>, O<sub>2</sub>, N<sub>2</sub>
 
*Gases Available: CH<sub>4</sub>, H<sub>2</sub>, Ar, Cl<sub>2</sub>, BCl<sub>3</sub>, SF<sub>6</sub>, SiCl<sub>4</sub>, O<sub>2</sub>, N<sub>2</sub>
 +
*ICP Power (max): 3000 W
 +
*RF Power (max): 600 W
 
*He-back-side cooling
 
*He-back-side cooling
 
*100mm wafer held down with ceramic clamp., single-load
 
*100mm wafer held down with ceramic clamp., single-load
Line 44: Line 44:
  
 
*Laser monitoring with camera and etch simulation software: [[Laser Etch Monitoring|Intellemetrics LEP 500]]
 
*Laser monitoring with camera and etch simulation software: [[Laser Etch Monitoring|Intellemetrics LEP 500]]
 +
*Optical Emission Spectroscopy (Ocean Optics) for endpoint detection of chamber cleans & etches - integrated into Oxford software
  
 
==Documentation==
 
==Documentation==

Revision as of 10:59, 29 September 2021

Oxford ICP Etcher (PlasmaPro 100 Cobra)
OxfordPlasmaPro.jpg
Tool Type Dry Etch
Location Bay 2
Supervisor Tony Bosch
Supervisor Phone (805) 893-3486
Supervisor E-Mail bosch@ece.ucsb.edu
Description ICP Etches for III-V/ALE
Manufacturer Oxford Instruments
Model PlasmaPro 100 Cobra 300
Materials InP, GaAs, GaN, Silicon ALE
Dry Etch Recipes


About

The Oxford PlasmaPro 100 Cobra 300 is intended for etching InP-based, GaAs-baased and GaN-based epitaxies, in addition to Atomic Layer Etching (ALE) processes. The system has a load lock, wide temperature range with rapid heating/cooling, Inductively Coupled Plasma (ICP) coil and a capactively coupled substrate HF (13.56MHz) The fixturing is configured for 4" diameter Si wafers and uses a clamp to hold the sample on the RF chuck. Small pieces may be placed on Silicon carrier wafers, with or without mounting adhesive. Helium back-side cooling is used to keep the sample cool during the etch, but pieces do heat up when placed on carriers.

The in-situ laser monitor installed on this system allows for repeatable etches and endpoint detection via continuous optical monitoring of the wafer reflectivity in a user-determined location, through a porthole on the chamber. The system also has an in situ optical emission monitor for plasma spectroscopy, utilized for chamber clean endpoint detection.

Detailed Specifications

  • Temperature Range: –40°C to +400°C
  • Gases Available: CH4, H2, Ar, Cl2, BCl3, SF6, SiCl4, O2, N2
  • ICP Power (max): 3000 W
  • RF Power (max): 600 W
  • He-back-side cooling
  • 100mm wafer held down with ceramic clamp., single-load
    • Users may place pieces onto carrier wafer with or without adhesive. Standard recipes use no adhesive.
    • Pieces must be >7mm from edge of carrier to avoid wafer-clamping mechanism.
  • Windows-based Cortex software control of process and wafer handling
  • Allowed Materials:
    • InP-based epitaxies - qualified and ready
    • GaAs-baased epitaxies - discuss with staff
    • GaN-based epitaxies - discuss with staff
    • GaSb-based epitaxies - discuss with staff
    • Atomic Layer Etching on select materials - discuss with staff
  • Standard masking materials include:
    • SiO2
    • Si3N4
    • photoresist (at << 100°C).

Other materials can be exposed to the chamber only with staff approval.

  • Laser monitoring with camera and etch simulation software: Intellemetrics LEP 500
  • Optical Emission Spectroscopy (Ocean Optics) for endpoint detection of chamber cleans & etches - integrated into Oxford software

Documentation