Difference between revisions of "Lift-Off with DUV Imaging + PMGI Underlayer"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
m (convert process to table, more like a "traveler")
(separated Process Limits into separate section)
Line 1: Line 1:
 
Developed by Demis D. John, ~2017-2019
 
Developed by Demis D. John, ~2017-2019
 
This process is intended for Deep-UV Exposure on the ASML DUV Stepper. PMGI is used as the underlayer, which is exposed at the same time as the imaging resist.
 
This process is intended for Deep-UV Exposure on the ASML DUV Stepper. PMGI is used as the underlayer, which is exposed at the same time as the imaging resist.
 
This process uses PMGI SF-5 @ 4krpm, which spins to about 130nm (maximum metal thickness = 130nm/3 ≈ 45nm) and undercuts laterally about 130-140nm (minimum gap between metals = (2 x 140nm) + 30nm ≈ 350nm).
 
   
 
Adjust spin speed or switch to a different PMGI formulation to tailor the underlayer thickness to your desired metal thickness.
 
Adjust spin speed or switch to a different PMGI formulation to tailor the underlayer thickness to your desired metal thickness.
Line 18: Line 16:
   
 
*Make sure to use a vertical evaporation - EBeam4 or EBeam1. EBeam3 is not vertical.
 
*Make sure to use a vertical evaporation - EBeam4 or EBeam1. EBeam3 is not vertical.
  +
  +
=== Process Limits ===
  +
This process is written with PMGI SF-5 @ 4krpm, which spins to about 130nm
  +
* <u>Maximum metal thickness</u> = 130nm/3 ≈ '''45nm'''
  +
* Undercuts laterally about 130-140nm, so <u>minimum gap between metals</u> = (2 x 140nm) + ~30nm ≈ '''350nm'''
  +
Adjust the PMGI layer thickness to adjust these process limits.
   
 
==Suggested Process for Liftoff==
 
==Suggested Process for Liftoff==

Revision as of 17:59, 24 June 2020

Developed by Demis D. John, ~2017-2019

This process is intended for Deep-UV Exposure on the ASML DUV Stepper. PMGI is used as the underlayer, which is exposed at the same time as the imaging resist.

Adjust spin speed or switch to a different PMGI formulation to tailor the underlayer thickness to your desired metal thickness.

Tips

Please see our [Bi-Layer Lift-Off Tutorial] to understand the limitations and requirements for good lift-off.

  • Remember that gaps in between adjacent metal traces can lift-off during the develop!
  • Need underlayer thickness 2x to 3x the desired metal thickness.
  • The underlayer will develop laterally from both sides which can lift-off the imaging resist, so:
    • Minimum gap between adjacent metals should then be: less than 0.5 * underlayer thickness
    • Want at least 30-50nm of underlayer width left to support the imaging resist, so the PR doesn't fall over/collapse.
  • Make sure to use a vertical evaporation - EBeam4 or EBeam1. EBeam3 is not vertical.

Process Limits

This process is written with PMGI SF-5 @ 4krpm, which spins to about 130nm

  • Maximum metal thickness = 130nm/3 ≈ 45nm
  • Undercuts laterally about 130-140nm, so minimum gap between metals = (2 x 140nm) + ~30nm ≈ 350nm

Adjust the PMGI layer thickness to adjust these process limits.

Suggested Process for Liftoff

De-H2O Bake (eg. 220°C, 1m+)

- or - Technics O2 ash (300mT, 100W), 15sec

Oxygen ash is faster and more effective
Spin PMGI SF-5 @ 4krpm (rcp 7)
POLOS underside clean: 2000rpm, ACE/ISO/N2
PMGI-Bake: 220°C, 3min (BrewerSci lift-pin hotplate). Can just place wafer directly onto hotplate surface.

Optionally can use lift-pins & Recipe "00 220deg, 3min Vac";

--> Enable Vacuum with overhead valve, only for wafer landing and then turn off to prevent sliding.

Spin UV-6-0.8 @ rcp 6 (3.5krpm)
POLOS underside clean: 2000rpm, ACE/ISO/N2
Soft-Bake = 135°C, 1min (builtin hotplate)
(check underside for particulates)
ASML Exposure
Default: Exp = 37.5mJ // foc = –0.10 Try this exposure dose but might need to do a FocArray (smaller Image Distribution eg. 5x5) to find proper exposure.
PEB = 135°C, 1.5min (built-in hotplate)
Dev (300MiF) = 50sec (CRITICAL time) Use cassette & H2O rinse dish prepared

--> not stirring, instead very slow/gentle lift/drop at various angles or gentle swish just to mix developer slightly.

DI rinse Very Gently! Dunk into prepared DI dish, and dump/fill gently 2x. Don’t allow direct hard water to hit PR surface.
PEii Technics O2: 30sec (100W/300mT) May increase feature size openings by ~50nm.
Metal Evaporation

(insert steps)

EBeam#4 or EBeam#1 with vertical/4-inch holders (not planetary)
Lift-off in NMP Facing down or vertical.

Data

• For PMGI Baked at 220°C for 3min, the PMGI dissolves in AZ 300MIF at a rate of approximately ~600nm/min.

Images / Examples

SEM of PR cross-section
SEM of holes in PR with above process. Lateral measurement of undercut is only approximate because circle may not have been cut through the exact diameter. (Demis D. John, 2017)