Difference between revisions of "Stepper 3 (ASML DUV)"

From UCSB Nanofab Wiki
Jump to: navigation, search
m (moved/updated "service provider")
(34 intermediate revisions by 3 users not shown)
Line 2: Line 2:
 
|picture=ASML.jpg
 
|picture=ASML.jpg
 
|type = Lithography
 
|type = Lithography
|super= Adam Abrahamsen
+
|super= Demis D. John
  +
|location=Bay 7
|phone=(805)839-3918x213
 
|location=Bay 3
 
|email=abrahamsen@ece.ucsb.edu
 
 
|description = ASML PAS 5500/300 DUV Stepper
 
|description = ASML PAS 5500/300 DUV Stepper
|manufacturer = ASML
+
|manufacturer = [http://www.asml.com ASML]
  +
|ToolType = Lithography
  +
|recipe = Lithography
 
|materials =
 
|materials =
 
|toolid=51
 
|toolid=51
 
}}
 
}}
   
= About =
+
=About=
   
The ASML 5500 stepper is a 248nm DUV stepper for imaging dense features down to below 200nm and isolated line structures down to below 150nm. The system has a variable NA system and has a field image size of 21 x 21mm for 0.63 NA and a field size of 22mm x 27mm for 0.4 to 0.57 NA. Overlay accuracy is better than 30nm. The system is configured for 4” wafers and pieces down to 14mm in size can be exposed using a 4” wafer as a carrier.
+
The ASML 5500 stepper is a 248nm DUV stepper for imaging dense features down to below 200nm and isolated line structures down to below 150nm. Overlay accuracy is better than 30nm.
   
  +
The system is configured for 4” wafers and, with staff support, mounted pieces down to 14mm in size can be exposed using a 4” wafer as a carrier. The system is designed for high throughput, so shooting multiple 4" wafers is extremely fast. Additionally, exposure jobs are highly programmable, allowing for very flexible exposures of multiple aligned patterns from multiple masks in a single session, allowing for process optimization of large vs. small features in a single lithography.
= Process Information =
 
   
  +
The full field useable exposure area is limited to the intersection of a 31mm diameter circle and a rectangle of dimensions 22mm x 27mm. See the [[ASML 5500 Mask Making Guidelines|Mask Making Guidelines page]] for more info on exposure field sizes and how to order your mask plates.
*[http://signupmonkey.ece.ucsb.edu/wiki/index.php/Lithography_Recipes Process Page]
 
   
  +
Resists Used (see [https://www.nanotech.ucsb.edu/wiki/index.php/Lithography_Recipes#Photolithography_Recipes PhotoLith. Recipes] for full process info):
=Service Provider=
 
  +
* [http://www.asml.com ASML]
 
  +
*UV210-0.3 - Positive: 300nm nominal thickness
  +
*UV6-0.8 - Positive: 800nm nominal thickness
  +
*UV26-2.5 - Positive: 2.5um nominal thickness
  +
*UVN2300-0.5 - Negative: 500nm nominal thickness
  +
  +
*DUV42P-6/DS-K101 - Bottom Anti-Reflective Coatings “BARC”
  +
*PMGI/LOL1000/LOL2000 - Underlayers
  +
  +
AZ300MIF Developer for all processes
  +
  +
=Process Information=
  +
  +
*[https://www.nanotech.ucsb.edu/wiki/index.php/Lithography_Recipes#Photolithography_Recipes '''Process Recipes Page'''] '''> "Stepper 3"''' - ''Established recipes and corresponding linewidths, photoresists etc.''
  +
*Sample size: 100 mm wafers with SEMI std. major flat
  +
**''Piece-parts process is possible but difficult - contact staff for info''
  +
*Alignment Accuracy: < 50 nm
  +
*Minimum Feature Size: ≤150 nm isolated lines, ≤200 nm dense patterns
  +
*Maximum Wafer Bow: approx. 100 µm. (4-inch diam.)
  +
**''Near this value, and the job may fail or lose the wafer inside the machine due to wafer vacuum error.''
  +
*[[ASML Stepper 3 - UCSB Test Reticles|UCSB Test Reticles - Alignment Markers, Resolution Testing etc.]]
   
 
=Operating Procedures=
 
=Operating Procedures=
  +
* [[media:GCA-200-Training-Manual.pdf|Training Manual]]
 
  +
*[[ASML Stepper 3 Standard Operating Procedure|Standard Operating Procedures]]
  +
**''Exposing wafers, loading reticles, focus/exposure matrix''
  +
*[//www.nanotech.ucsb.edu/wiki/images/f/f7/ASML_Job_Set-Up_Guide_v2.pdf Job Programming - Full]
  +
*[//www.nanotech.ucsb.edu/wiki/images/c/cd/ASML_Job_Set-Up_Guide_simple_v1.pdf Job Programming- Simplified -Full Wafers]
  +
*[[ASML 5500 Mask Making Guidelines|ASML 5500 Mask Making Guidelines]]
  +
  +
===Troubleshooting and Recovery===
  +
  +
*[[ASML Stepper 3 Error Recovery, Troubleshooting and Calibration|Error Recovery, Troubleshooting and Calibration]]
  +
**''Common errors/System Warnings, Wafer Handler Reset, System Calibration Verification''
  +
  +
*[[ASML 5500: Recovering from an Error|ASML 5500: Recovering from an Error/Wafer Retrieval]]
  +
**''How to abort the job and recover your wafer.''
  +
  +
=Service Provider=
  +
  +
*[http://www.asml.com ASML] - ASML performs quarterly periodic maintenance and provides on-demand support.

Revision as of 16:34, 21 November 2019

Stepper 3 (ASML DUV)
ASML.jpg
Tool Type Lithography
Location Bay 7
Supervisor Demis D. John
Supervisor Phone (805) 893-5934
Supervisor E-Mail demis@ucsb.edu
Description ASML PAS 5500/300 DUV Stepper
Manufacturer ASML
Lithography Recipes
Sign up for this tool



About

The ASML 5500 stepper is a 248nm DUV stepper for imaging dense features down to below 200nm and isolated line structures down to below 150nm. Overlay accuracy is better than 30nm.

The system is configured for 4” wafers and, with staff support, mounted pieces down to 14mm in size can be exposed using a 4” wafer as a carrier. The system is designed for high throughput, so shooting multiple 4" wafers is extremely fast. Additionally, exposure jobs are highly programmable, allowing for very flexible exposures of multiple aligned patterns from multiple masks in a single session, allowing for process optimization of large vs. small features in a single lithography.

The full field useable exposure area is limited to the intersection of a 31mm diameter circle and a rectangle of dimensions 22mm x 27mm. See the Mask Making Guidelines page for more info on exposure field sizes and how to order your mask plates.

Resists Used (see PhotoLith. Recipes for full process info):

  • UV210-0.3 - Positive: 300nm nominal thickness
  • UV6-0.8 - Positive: 800nm nominal thickness
  • UV26-2.5 - Positive: 2.5um nominal thickness
  • UVN2300-0.5 - Negative: 500nm nominal thickness
  • DUV42P-6/DS-K101 - Bottom Anti-Reflective Coatings “BARC”
  • PMGI/LOL1000/LOL2000 - Underlayers

AZ300MIF Developer for all processes

Process Information

  • Process Recipes Page > "Stepper 3" - Established recipes and corresponding linewidths, photoresists etc.
  • Sample size: 100 mm wafers with SEMI std. major flat
    • Piece-parts process is possible but difficult - contact staff for info
  • Alignment Accuracy: < 50 nm
  • Minimum Feature Size: ≤150 nm isolated lines, ≤200 nm dense patterns
  • Maximum Wafer Bow: approx. 100 µm. (4-inch diam.)
    • Near this value, and the job may fail or lose the wafer inside the machine due to wafer vacuum error.
  • UCSB Test Reticles - Alignment Markers, Resolution Testing etc.

Operating Procedures

Troubleshooting and Recovery

Service Provider

  • ASML - ASML performs quarterly periodic maintenance and provides on-demand support.