Difference between revisions of "PECVD 1 (PlasmaTherm 790)"
(Created page with "{{tool|{{PAGENAME}} |picture=e-beam1.jpg |type = Vacuum Deposition |super= Brian Lingg |phone=(805)839-3918x210 |location=Bay 3 |email=lingg@ece.ucsb.edu |description = Four Pock…") |
|||
(32 intermediate revisions by 3 users not shown) | |||
Line 1: | Line 1: | ||
{{tool|{{PAGENAME}} |
{{tool|{{PAGENAME}} |
||
− | |picture= |
+ | |picture=PECVD1.jpg |
|type = Vacuum Deposition |
|type = Vacuum Deposition |
||
|super= Brian Lingg |
|super= Brian Lingg |
||
Line 6: | Line 6: | ||
|location=Bay 3 |
|location=Bay 3 |
||
|email=lingg@ece.ucsb.edu |
|email=lingg@ece.ucsb.edu |
||
− | |description = |
+ | |description = PECVD Plasma Therm 790 For Oxides And Nitrides |
− | |manufacturer = |
+ | |manufacturer = Plasma-Therm |
|materials = |
|materials = |
||
+ | |toolid=16 |
||
}} __TOC__ |
}} __TOC__ |
||
− | |||
= About = |
= About = |
||
Line 27: | Line 27: | ||
*Full computer operation |
*Full computer operation |
||
*Standard recipes for a variety of film thicknesses |
*Standard recipes for a variety of film thicknesses |
||
+ | |||
+ | =Documentation= |
||
+ | *[[media:PECVD1 Operating Instructions.pdf|Operating Instuctions]] |
Revision as of 11:46, 9 January 2015
|
About
This is a Plasma-Therm model 790 plasma enhanced chemical vapor deposition system for depositing SiO2, Si3N4, or SiOxNy dielectric films. The system uses a capacitively-coupled 13.56 MHz source excitation to produce the plasma between two parallel aluminum plates. The gas is injected over the sample through a 6” diameter showerhead. The samples are placed on the system anode (to minimize ion damage) which is heated to 250-350°C. SiO2 is produced from SiH4/He 2%/98% and N2O at 250°C. The typical deposition rate is 400 A/min. at 300 mT pressure. The typical BOE etch rate of this oxide is about 400 nm/min. Si3N4 is produced from SiH4/He 2%/98% and NH3 at 250°C or 350°C. The more dense films are produced at 350°C. The stress of the nitride can be altered by adjusting the N2:He ratio of the deposition. CF4/O2 plasmas are used to clean the chamber between depositions.
These films are typically used for capacitor dielectrics, chemical passivation layers, electrical insulators, reactive ion etching masks, and optical anti-reflective coatings. The system is fully programmable with windows-based software and has a wide array of pre-defined thicknesses. Custom programs for dielectric stacks or different process parameters can be written and saved.
Detailed Specifications
- Gases used: NH3, N2O, 2%SiH4/He, N2,CF4 and O2
- ~ 10mT ultimate chamber pressure
- 13.56 Mhz excitation freq.
- Sample size: pieces to 6” wafers
- Automatic tuning network
- RF Power control
- Full computer operation
- Standard recipes for a variety of film thicknesses