Difference between revisions of "E-Beam 3 (Temescal)"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
m (→‎About: Removed the word rotated.)
(29 intermediate revisions by 2 users not shown)
Line 13: Line 13:
 
= About =
 
= About =
   
This electron-beam evaporation system is the work-horse of the lab for metal deposition. The system has the unique feature of a home-built load-lock system that allows very quick cycle time for evaporation (as low as 20 minutes total time). The system also has two 4-pocket e-beam sources and an Inficon IC/5 deposition controller that allows for co-deposition of certain metals. The front gun contains metals Ti, Pt, Ni, Au and the back gun contains metals Pd, Al, Ag, Ge. These metals stay under high vacuum at all times, except during maintenance, to maintain source purity. One wafer up to 4” diameter or multiple pieces can be placed into this system for evaporation. There is also a special fixture that can be inserted for angling and/or rotating the sample during deposition. This system is used for n-type ohmic contact metalization to compound semiconductors, Schottky contacts to semiconductors, bond pads, and other general metalizations. The maximum deposition thickness during a run is limited to 1 micron.
+
This electron-beam evaporation system is the work-horse of the lab for metal deposition. The system has the unique feature of a home-built load-lock system that allows very quick cycle time for evaporation (as low as 20 minutes total time). The system also has two 4-pocket e-beam sources and an Inficon IC/5 deposition controller that allows for co-deposition of certain metals. The front gun contains metals Ti, Pt, Ni, Au and the back gun contains metals Pd, Al, Ag, Ge. These metals stay under high vacuum at all times, except during maintenance, to maintain source purity. One wafer up to 4” diameter or multiple pieces can be placed into this system for evaporation. There is also a special fixture that can be inserted for angling the sample during deposition. This system is used for n-type ohmic contact metalization to compound semiconductors, Schottky contacts to semiconductors, bond pads, and other general metalizations. The maximum deposition thickness during a run is limited to 1 micron.
   
 
= Detailed Specifications =
 
= Detailed Specifications =
Line 22: Line 22:
 
*Load-lock for quick turn-around
 
*Load-lock for quick turn-around
 
*Automatic vacuum sequencing
 
*Automatic vacuum sequencing
  +
  +
 
*Temescal Super Sweep e-beam sweep control
 
*Temescal Super Sweep e-beam sweep control
 
*Inficon IC/5 programmable crystal thickness monitoring system
 
*Inficon IC/5 programmable crystal thickness monitoring system
Line 30: Line 32:
 
 
 
=Documentation=
 
=Documentation=
  +
*{{file|E-BEAM-3-OPERATING- INSTRUCTIONS-1.PDF|E-Beam#3 operating instructions}}
*[[media*[[media:Web_Operational_Procedure_EB4_Rev_3.pdf|Operating Procedures]]
 
|Operating Procedures]]
 
   
 
= Materials Table =
 
= Materials Table =

Revision as of 15:24, 7 November 2018

E-Beam 3 (Temescal)
E-beam3.jpg
Tool Type Vacuum Deposition
Location Bay 3
Supervisor Don Freeborn
Supervisor Phone (805) 893-7975
Supervisor E-Mail dfreeborn@ece.ucsb.edu
Description Load Locked Metal Evaporator Dual Gun
Manufacturer Temescal
Vacuum Deposition Recipes
Sign up for this tool


About

This electron-beam evaporation system is the work-horse of the lab for metal deposition. The system has the unique feature of a home-built load-lock system that allows very quick cycle time for evaporation (as low as 20 minutes total time). The system also has two 4-pocket e-beam sources and an Inficon IC/5 deposition controller that allows for co-deposition of certain metals. The front gun contains metals Ti, Pt, Ni, Au and the back gun contains metals Pd, Al, Ag, Ge. These metals stay under high vacuum at all times, except during maintenance, to maintain source purity. One wafer up to 4” diameter or multiple pieces can be placed into this system for evaporation. There is also a special fixture that can be inserted for angling the sample during deposition. This system is used for n-type ohmic contact metalization to compound semiconductors, Schottky contacts to semiconductors, bond pads, and other general metalizations. The maximum deposition thickness during a run is limited to 1 micron.

Detailed Specifications

  • Temescal CV-6S 10kV power supply
  • 2-Temescal 4-pocket series 260 e-beam sources
  • Turbo-pumped system with ~ 5e-8 ultimate base pressure
  • Load-lock for quick turn-around
  • Automatic vacuum sequencing


  • Temescal Super Sweep e-beam sweep control
  • Inficon IC/5 programmable crystal thickness monitoring system
  • Sample size: 1 wafer up to 4” diameter
  • Metals:
    • Front Gun: Ti, Pt, Ni, Au
    • Rear Gun: Pd, Al, Ag, Ge

Documentation

Materials Table

For the materials tables, please visit the E-Beam Recipe Page.