Line 42: Line 42:
 
|0.336
 
|0.336
 
|140.0
 
|140.0
  +
|OK
|Tony, could you please check this?
 
 
|-
 
|-
 
|GeO<sub>2</sub>
 
|GeO<sub>2</sub>
Line 60: Line 60:
 
|0.411
 
|0.411
 
|157.6
 
|157.6
  +
|OK
|Check this
 
 
|-
 
|-
 
|Si
 
|Si

Revision as of 12:11, 29 August 2012

E-Beam 2 (Custom)
E-beam2.jpg
Tool Type Vacuum Deposition
Location Bay 3
Supervisor Tony Bosch
Supervisor Phone (805) 893-3486
Supervisor E-Mail bosch@ece.ucsb.edu
Description Electron-Beam Evaporation System
Manufacturer Temescal
Vacuum Deposition Recipes
Sign up for this tool


About

This electron-beam evaporation system is used for the controlled deposition of thin dielectric films. The films are evaporated from a wide variety of solid sources. The most common dielectrics deposited are: SiO2, SiO, TiO2, Ta2O5, SrF2. Other materials may be evaporated upon request. Oxygen gas can be bled into the system during deposition to try to maintain the stoichiometry during deposition. Fixturing for heating the substrate can also be used. A crystal thickness monitor is used to control the deposition thickness. The dielectrics deposited by this system are typically used for optical coatings (anti-reflective and highly reflective multiple layer stacks), electrical insulators, and reactive ion etching masks. Samples up to ~ 3” x 3” can be placed into this system for evaporation. Typical deposition rates are several Angstroms/second.

Detailed Specifications

  • Temescal 10kV power supply
  • Temescal 4-pocket series 260 e-beam source
  • Electron beam controller with sweep rate and amplitude control
  • Cryo-pumped system with ~ 1e-7 ultimate base pressure
  • Automatic vacuum sequencing
  • Crystal thickness monitoring
  • Sample size: up to 3” x 3” pieces
  • Oxygen bleed for maintaining oxide stoichiometry


Materials Table