AUTOSTEP 200 CHUCKS
Chuck Name / Wafer Size Substrate Thickness Chuck Thickness
1/4 of 2" 500 um 11.646 mm
2" 500 um 11.650 mm
100mm (or 4") 500 um 11.630 mm
1/4 of 2" 635 um 11.536 mm
76.2mm (or 3") 635 um 11.570 mm
Al 100 mm 635 um 11.570 mm
Black Multi 720 um 11.440 mm
TARGET THICKNESS: 12.150mm +/- 0.1mm( 100um)
Target Thickness = Chuck Thickness + Wafer Thickness
Limits on Height/Focus Display ( +/-6V), Ideal ≈ 0V
NOTE: DO NOT OVER TIGHTEN SCREWS!