Unaxis VLR Etch - Process Control Data

From UCSB Nanofab Wiki
Revision as of 11:08, 27 April 2022 by Pakala (talk | contribs) (merged the two sets of ICP cal data)
Jump to navigation Jump to search
UnderConstruction.jpg

Work In Progress

This article is still under construction. It may contain factual errors. Content is subject to change.


Data - InP Ridge Etch (Unaxis VLR)

PECVD SiO2 hardmask, patterned on Stepper #2 (AutoStep 200) & Panasonic ICP #1

InP Ridge Etch: 200°C, 1.4mT, 800W/125W, Cl2=6.3, H2=12.7, Ar=2.0 sccm, time=1min30sec (90sec)

Sample Size: 1x1cm epi-grade InP, ~30-40% SiO2 masking (NingC's pattern). Silicon carrier, no adhesive.

Conditioning: Prior to the etch, do O2 clean 15 minutes, then, chamber coating with the same recipe on 1/4-2" InP on Silicon carrier for 15 minutes.

Date Sample# Etch Rate (nm/min) Etch Selectivity (InP/SiO2) Comments SEM Images
3/30/22 NP_Unaxis_02 1413 14.6 [1] [2]
3/9/22 NP_Unaxis_01 1297 15.3 [1] [2]
11/8/2021 InP#2102 1.24 13.8 [1][2]
2/3/2021 InP#2101 1.30 16 [3][4]
8/30/2020 InP#2001 1.11 10.4 [5]
1/31/2019 InP#1901 0.88 9.7 [6][7]
12/10/2018 InP#1809 1.01 11.4 [8]
10/3/2018 InP#1808 1.01 13.7 [9]
8/7/2018 InP#1807 0.81 8.0 [10]
5/22/2018 InP#1806 0.88 8.4 [11]
4/26/2018 InP#1805 1.29 13.6 [12]
4/10/2018 InP#1804 1.12 12.8 [13]
4/5/2018 InP#1803 1.05 11.9 [14]
3/1/2018 InP#1802 0.96 9 [15]
1/2/2018 InP#1801 1.44 14.3 [16]
12/7/2017 InP#1714 0.96 10.4 [17]
11/21/2017 InP#1713 1.04 12.1 [18]
10/23/2017 InP#1712 1.11 13.1 [19]
10/11/2017 InP#1711 1 11 [20]
8/28/2017 InP#1710 1 11.7 [21]
8/16/2017 InP#1709 0.76 8 [22]
7/6/2017 InP#1708 0.98 12.1 [23]
5/19/2017 InP#1707 0.82 9.9 [24]
5/4/2017 InP#1706 0.84 11 [25]
4/20/2017 inP#1705 0.88 10.2 [26]
3/21/2017 InP#1704 1.01 11.3 [27]
2/21/2017 InP#1703 0.91 11.3 [28]
2/7/2017 InP#1702 0.75 7.7 [29]
1/23/2017 InP#1701 0.93 9.4 [30]
12/15/2016 InP#1615 0.91 9.3 [31]
12/1/2016 InP#1614 0.96 12.1 [32]
10/4/2016 InP#1613 0.92 8.9 [33]