(Created page with "{{tool|{{PAGENAME}} |picture=WaferThermco.jpg |type = Thermal Processing |super= Mike Silva |phone=(805)839-3918x219 |location=Bay 4 |email=silva@ece.ucsb.edu |description = Wafe…") |
|||
(One intermediate revision by one other user not shown) | |||
Line 2: | Line 2: | ||
|picture=WaferThermco.jpg |
|picture=WaferThermco.jpg |
||
|type = Thermal Processing |
|type = Thermal Processing |
||
− | |super= |
+ | |super= Tony Bosch |
|phone=(805)839-3918x219 |
|phone=(805)839-3918x219 |
||
|location=Bay 4 |
|location=Bay 4 |
||
Line 9: | Line 9: | ||
|manufacturer = Thermco |
|manufacturer = Thermco |
||
|materials = |
|materials = |
||
+ | |toolid=43 |
||
}} |
}} |
||
= About = |
= About = |
Revision as of 10:47, 15 April 2020
|
About
This wafer fusion furnace is a custom built unit designed for the fusion of different material types (such InP and Si) to create new heterostructures for novel electronic and optoelectronic devices. Custom made bonding fixtures are loaded into the furnace through a purged glove-box that is oxygen free to below 10 ppm with a dew point near -80°C. The furnace is nitrogen purged and can operate up from 200°C to 1200°C. Many different materials are allowed and maximum temperature is dependent on the materials themselves, to avoid contamination of the furnace.
Detailed Specifications
- Temperatures 200°C to 1200°C
- 4 inch diameter furnace
- Nitrogen ambient at atmospheric pressure
- Oxygen control in glove box to less than 10 ppm
- Dew point near -80°C
- Custom bonding fixtures of various materials allowed