Difference between revisions of "Test Data of etching SiO2 with CHF3/CF4-ICP1"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
(Text replacement - "www.nanotech.ucsb.edu/wiki/" to "wiki.nanotech.ucsb.edu/wiki/")
(comments on sidewall angle)
Line 14: Line 14:
 
|110
 
|110
 
|1.35
 
|1.35
  +
|see SEM →
|
 
 
|[https://wiki.nanotech.ucsb.edu/wiki/images/b/b1/I11901.pdf]
 
|[https://wiki.nanotech.ucsb.edu/wiki/images/b/b1/I11901.pdf]
 
|-
 
|-
Line 28: Line 28:
 
|78.0
 
|78.0
 
|1.06
 
|1.06
  +
|unusual - two regions
|
 
 
|[https://wiki.nanotech.ucsb.edu/wiki/images/b/bb/I1200107.pdf]
 
|[https://wiki.nanotech.ucsb.edu/wiki/images/b/bb/I1200107.pdf]
 
|-
 
|-
| colspan="6" |Rate on 1/13/2020 is low by ~20%, so chamber was wet-cleaned on 1/21/2020. Data below for 1/23 shows rate returned to "normal".
+
| colspan="6" |''Rate on 1/13/2020 is low by ~20%, so chamber was wet-cleaned on 1/21/2020. Data below for 1/23 shows rate returned to "normal".''
 
|-
 
|-
 
|1/23/2020
 
|1/23/2020
Line 47: Line 47:
 
|[https://wiki.nanotech.ucsb.edu/wiki/images/e/ec/I1200301.pdf]
 
|[https://wiki.nanotech.ucsb.edu/wiki/images/e/ec/I1200301.pdf]
 
|-
 
|-
| colspan="6" |Sidewall profile on 2/28/2020 etches is slanted, chamber will be wet-cleaned on 3/2/2020 and re-tested.
+
| colspan="6" |''Sidewall profile on 2/28/2020 etches is slanted, chamber will be wet-cleaned on 3/2/2020 and re-tested. Data below shows etch returned to "normal".''
 
|-
 
|-
 
|3/3/2020
 
|3/3/2020

Revision as of 09:10, 23 April 2020

ICP#1: 0.5Pa, 50/900W, CHF3/CF4=10/30 sccm, time=210 sec
Date Sample# Etch Rate (nm/min) Etch Selectivity (SiO2/PR) Averaged Sidewall Angle (o) SEM Images
1/28/2019 I11901 110 1.35 see SEM → [1]
5/29/2019 I11903 105 1.41 [2]
1/13/2020 I12001 78.0 1.06 unusual - two regions [3]
Rate on 1/13/2020 is low by ~20%, so chamber was wet-cleaned on 1/21/2020. Data below for 1/23 shows rate returned to "normal".
1/23/2020 I12002 109 1.16 [4]
2/28/2020 I12003 119 1.17 56.6 [5]
Sidewall profile on 2/28/2020 etches is slanted, chamber will be wet-cleaned on 3/2/2020 and re-tested. Data below shows etch returned to "normal".
3/3/2020 I12004 110 1.05 [6]