Difference between revisions of "Template:Announcements"
(remove asml) |
(delete SLR turbo fail, AMSL maint.) |
||
Line 9: | Line 9: | ||
<!------------- Announcements ----------------> |
<!------------- Announcements ----------------> |
||
− | ===== |
+ | ===== ASML Maintenance ===== |
+ | ASML is here performing annual maintenance through Wednesday at least, possibly through Friday. |
||
− | The SLR fluorine etcher is down, the turbo has failed. |
||
− | // [[User:John d|John d]] |
+ | // [[User:John d|John d]] 11:15, 17 July 2018 (PDT) |
+ | |||
===== PECVD#1: Software Upgrade ===== |
===== PECVD#1: Software Upgrade ===== |
Revision as of 11:15, 17 July 2018
UCSB NanoFab Announcements
ASML Maintenance
ASML is here performing annual maintenance through Wednesday at least, possibly through Friday. // John d 11:15, 17 July 2018 (PDT)
PECVD#1: Software Upgrade
We are going to upgrade the computer and software on PECVD#1 in 2–3 weeks. The tool will be down for three to five days. Once the upgrade is complete, everyone will need to be retrained on the new software. You should start planning accordingly, if you use this tool. We will send out an email when we know the exact dates. // John d 15:05, 20 June 2018 (PDT)
RIE#5: SiCl4 Issue
We have a flow rate problem on the SiCl4 gas line. You can only flow the SiCl4 for 5 minutes and then you have to wait 5 minutes before you flow it again. So in order to to do a 10 minute Etch you will have to do it in three steps flow 5 mi, wait 5mi, then flow 5min. // John d 17:22, 15 June 2018 (PDT)