Difference between revisions of "Sputter 4 (AJA ATC 2200-V)"
(9 intermediate revisions by 5 users not shown) | |||
Line 2: | Line 2: | ||
|picture=Sputter4.jpg |
|picture=Sputter4.jpg |
||
|type = Vacuum Deposition |
|type = Vacuum Deposition |
||
− | |super= |
+ | |super= Tony Bosch |
|location=Bay 3 |
|location=Bay 3 |
||
|description =Seven-Target DC/RF Magnetron Sputtering System |
|description =Seven-Target DC/RF Magnetron Sputtering System |
||
Line 10: | Line 10: | ||
}} |
}} |
||
=About= |
=About= |
||
− | The Seven-Target DC/RF Sputtering System, built by AJA International uses planar magnetron sources. The sources are contained in tiltable sputter gun modules that allow for maintaining uniformity control at various sample heights. Cross contamination between sources is minimized by using a chimney configuration with very narrow source shutter gaps. Uniformity better than 2% is achieved for various sample heights. 2 DC sources and 1 RF sources allow for co-deposition of materials. Other materials, such as ITO, Si, Al, Zr, etc. can be reactively RF sputtered in an O<sub>2</sub> or N<sub>2</sub> environment to produce metal-oxides or nitrides. The deposition chamber is |
+ | The Seven-Target DC/RF Sputtering System, built by AJA International uses planar magnetron sources. The sources are contained in tiltable sputter gun modules that allow for maintaining uniformity control at various sample heights. Cross contamination between sources is minimized by using a chimney configuration with very narrow source shutter gaps. Uniformity better than 2% is achieved for various sample heights. 2 DC sources and 1 RF sources allow for co-deposition of materials. Other materials, such as ITO, Si, Al, Zr, etc. can be reactively RF sputtered in an O<sub>2</sub> or N<sub>2</sub> environment to produce metal-oxides or nitrides. The deposition chamber is load-locked, with automatic wafer transfer, providing for fast substrate transfer and consistent, low base pressure. Venting and evacuation are automated with a 1200 l/s turbo (capable of pumping O<sub>2</sub>) achieving < 5 E-8T ultimate pressure. A VAT gate valve is used for process pressure control independent of gas flow. Substrates are clip mounted onto 4", 6" or a generic carrier. Flow rates are controlled with standard mass flow controllers. Argon is used for the sputter gases, with N<sub>2</sub> and O<sub>2</sub> used for reactive sputtering. Gun power supplies include: 300W DC, 13.56 Mhz 300W RF and a 150W substrate RF supply for in-situ substrate biasing and pre-cleaning. Samples can be heated to 800°C. The system is recipe driven and computer controlled for reproducible results. Up to 6" round wafer sizes can be accomodated in the system. Magnetic materials are restricted from this system, but can be sputter deposited either Sputter 3 or 5. |
=Detailed Specifications= |
=Detailed Specifications= |
||
Line 19: | Line 19: | ||
*No magnetic materials |
*No magnetic materials |
||
*SiO<sub>2</sub>, SiN, ITO, AlN, and other metal-oxide/nitrides possible |
*SiO<sub>2</sub>, SiN, ITO, AlN, and other metal-oxide/nitrides possible |
||
− | *< |
+ | *< 5E-8T ultimate pressure (3 mT typical operating pressure), load-locked chamber |
− | *6" |
+ | *6" and 4" round sample holder |
*Gun Tilt and Sample height adjustment |
*Gun Tilt and Sample height adjustment |
||
*Deposition uniformity is ~ 1-2% over 4 " diameter |
*Deposition uniformity is ~ 1-2% over 4 " diameter |
||
− | *Up to |
+ | *Up to 800°C deposition temperature in O<sub>2</sub> environment |
− | *RF Biasing of sample during deposition |
+ | *RF Biasing of sample during deposition or as a preclean |
− | *Automatic wafer loading and recipe |
+ | *Automatic wafer loading and recipe driven process control. |
+ | *[https://signupmonkey.ece.ucsb.edu/cgi-bin/users/browse.cgi?tool_ID=21&B1=Show '''The Sputter #4 SignupMonkey Page'''] lists the currently installed sputter targets. |
||
+ | |||
+ | =Documentation= |
||
+ | |||
+ | ==Procedures== |
||
+ | |||
+ | ==Data== |
||
+ | [[Sputtering Recipes]] |
Revision as of 09:20, 20 April 2020
|
About
The Seven-Target DC/RF Sputtering System, built by AJA International uses planar magnetron sources. The sources are contained in tiltable sputter gun modules that allow for maintaining uniformity control at various sample heights. Cross contamination between sources is minimized by using a chimney configuration with very narrow source shutter gaps. Uniformity better than 2% is achieved for various sample heights. 2 DC sources and 1 RF sources allow for co-deposition of materials. Other materials, such as ITO, Si, Al, Zr, etc. can be reactively RF sputtered in an O2 or N2 environment to produce metal-oxides or nitrides. The deposition chamber is load-locked, with automatic wafer transfer, providing for fast substrate transfer and consistent, low base pressure. Venting and evacuation are automated with a 1200 l/s turbo (capable of pumping O2) achieving < 5 E-8T ultimate pressure. A VAT gate valve is used for process pressure control independent of gas flow. Substrates are clip mounted onto 4", 6" or a generic carrier. Flow rates are controlled with standard mass flow controllers. Argon is used for the sputter gases, with N2 and O2 used for reactive sputtering. Gun power supplies include: 300W DC, 13.56 Mhz 300W RF and a 150W substrate RF supply for in-situ substrate biasing and pre-cleaning. Samples can be heated to 800°C. The system is recipe driven and computer controlled for reproducible results. Up to 6" round wafer sizes can be accomodated in the system. Magnetic materials are restricted from this system, but can be sputter deposited either Sputter 3 or 5.
Detailed Specifications
- 7 target with DC or RF operation
- Reactive sputtering with N2 or O2 using RF
- Co-deposition of two materials: DC or RF
- No magnetic materials
- SiO2, SiN, ITO, AlN, and other metal-oxide/nitrides possible
- < 5E-8T ultimate pressure (3 mT typical operating pressure), load-locked chamber
- 6" and 4" round sample holder
- Gun Tilt and Sample height adjustment
- Deposition uniformity is ~ 1-2% over 4 " diameter
- Up to 800°C deposition temperature in O2 environment
- RF Biasing of sample during deposition or as a preclean
- Automatic wafer loading and recipe driven process control.
- The Sputter #4 SignupMonkey Page lists the currently installed sputter targets.