Difference between revisions of "RIE 3 (MRC)"
(updated SOP link) |
|||
(25 intermediate revisions by 5 users not shown) | |||
Line 1: | Line 1: | ||
{{tool|{{PAGENAME}} |
{{tool|{{PAGENAME}} |
||
− | |picture= |
+ | |picture=RIE3.jpg |
|type = Dry Etch |
|type = Dry Etch |
||
− | |super= |
+ | |super= Lee Sawyer |
− | |phone= 805-893- |
+ | |phone= 805-893-2123 |
− | |location=Bay |
+ | |location=Bay 2 |
− | |email= |
+ | |email=lee_sawyer@ucsb.edu |
− | |description = RIE # |
+ | |description = RIE #3 Fluorine-Based System MRC 51 |
|manufacturer = Materials Research Corporation (MRC) |
|manufacturer = Materials Research Corporation (MRC) |
||
|materials = |
|materials = |
||
+ | |toolid=26 |
||
}} |
}} |
||
⚫ | |||
⚫ | This is a Materials Research Corporation RIE-51 parallel plate, 13.56 MHz system used for etching with fluorine-containing gases (CF<sub>4</sub>, SF<sub>6</sub>, and CHF<sub>3</sub>). The system is used primarily for etching of Si, SiO<sub>2</sub>, and Si<sub>3</sub>N<sub>4</sub> films. Metals such as tungsten may also be etched. Tool features include: six inch diameter water cooled cathode/substrate platform, pyrex cylinder for plasma confinement and gas flow control, adjustable cathode-anode spacing, fixed DC bias or RF power control and a HeNe laser etch monitor with chart recorder. It is turbo pumped and has no loadlock. |
||
⚫ | |||
⚫ | CF<sub>4</sub>/O<sub>2</sub> and SF<sub>6</sub>/O<sub>2</sub> will etch Si, SiO<sub>2</sub> and Si<sub>3</sub>N<sub>4</sub> readily since free fluorine is readily liberated in the plasma. The oxygen (up to 40%) initially enhances the fluorine concentration resulting in a higher etch rate. The oxygen also minimizes polymer formation in CF<sub>4</sub>/O<sub>2</sub>. Too much Oxygen will compete for fluorine available, suppressing the etch rate. Argon can be added to increase the physical component of etching. The highest etch rates are achieved with SF<sub>6</sub> due to the ease of liberating fluorine compared with CF<sub>4</sub>. The relative etch rate decreases as one goes from Si to Si<sub>3</sub>N<sub>4</sub> to SiO<sub>2</sub>. CF4 and CHF3 can be used to selectivity etch SiO<sub>2</sub> over Si and resist due to increased polymer formation from the presence of hydrogen. This polymer layer is thicker on Si and resist than on SiO<sub>2</sub>. The trade-off is selectivity versus sidewall profile as the polymer will result in a tapered wall profile. Also, the polymer can be difficult to remove after etching. |
||
⚫ | This is a Materials Research Corporation RIE-51 parallel plate, 13.56 MHz system used for etching with fluorine-containing gases ( |
||
⚫ | The etches have good selectivity to many metals and semiconductors such as Ni, Al, Cr, Ti, GaAs, InP, and GaN. The system generally produces anisotropic etch profiles unless one goes into a purely chemical fluorine etch mode with higher pressure SF<sub>6</sub> processes. The system also has a strong loading effect so that larger substrates and open areas will require more feed gas and higher pressure to compensate. As a result, individual processes need to be characterized. |
||
⚫ | CF<sub>4</sub> |
||
⚫ | |||
⚫ | The etches have good selectivity to many metals and semiconductors such as Ni, Al, Cr, Ti, GaAs, InP, and GaN. The system generally produces anisotropic etch profiles unless one goes into a purely chemical fluorine etch mode with higher pressure |
||
⚫ | |||
⚫ | |||
⚫ | |||
⚫ | |||
+ | *Manual gas control |
||
+ | *Automatic pressure control |
||
⚫ | |||
+ | *Timer circuit for stopping the plasma |
||
⚫ | |||
⚫ | |||
+ | ==Documentation== |
||
⚫ | |||
+ | |||
⚫ | |||
+ | *[https://wiki.nanotech.ucsb.edu/w/images/2/2a/RIE_3_SOP_Rev_B.pdf RIE #3 Standard Operating Procedure] |
||
⚫ | |||
⚫ | |||
− | *HeNe and IR laser monitoring for endpoint |
||
⚫ | |||
− | *DC Bias or RF power control |
||
⚫ | |||
− | *Typical etch conditions for InGaAsP: |
||
− | **75 mT (CH<sub>4</sub>/H<sub>2</sub>/Ar : 4/20/10 sccm) |
||
− | **450v bias |
||
− | **~ 45 nm/min. etch rate |
Revision as of 17:34, 5 March 2020
|
About
This is a Materials Research Corporation RIE-51 parallel plate, 13.56 MHz system used for etching with fluorine-containing gases (CF4, SF6, and CHF3). The system is used primarily for etching of Si, SiO2, and Si3N4 films. Metals such as tungsten may also be etched. Tool features include: six inch diameter water cooled cathode/substrate platform, pyrex cylinder for plasma confinement and gas flow control, adjustable cathode-anode spacing, fixed DC bias or RF power control and a HeNe laser etch monitor with chart recorder. It is turbo pumped and has no loadlock.
CF4/O2 and SF6/O2 will etch Si, SiO2 and Si3N4 readily since free fluorine is readily liberated in the plasma. The oxygen (up to 40%) initially enhances the fluorine concentration resulting in a higher etch rate. The oxygen also minimizes polymer formation in CF4/O2. Too much Oxygen will compete for fluorine available, suppressing the etch rate. Argon can be added to increase the physical component of etching. The highest etch rates are achieved with SF6 due to the ease of liberating fluorine compared with CF4. The relative etch rate decreases as one goes from Si to Si3N4 to SiO2. CF4 and CHF3 can be used to selectivity etch SiO2 over Si and resist due to increased polymer formation from the presence of hydrogen. This polymer layer is thicker on Si and resist than on SiO2. The trade-off is selectivity versus sidewall profile as the polymer will result in a tapered wall profile. Also, the polymer can be difficult to remove after etching.
The etches have good selectivity to many metals and semiconductors such as Ni, Al, Cr, Ti, GaAs, InP, and GaN. The system generally produces anisotropic etch profiles unless one goes into a purely chemical fluorine etch mode with higher pressure SF6 processes. The system also has a strong loading effect so that larger substrates and open areas will require more feed gas and higher pressure to compensate. As a result, individual processes need to be characterized.
Detailed Specifications
- Etch gases include: CF4, CHF3, SF6, Ar, O2
- Low 1 E -6 ultimate chamber pressure
- 13.56 Mhz excitation frequency
- Manual gas control
- Automatic pressure control
- Manual RF tuning network
- Timer circuit for stopping the plasma
- Sample size limited to approximately 4 inches
- Masking materials include: Ni, photoresist (limited to low bias/power), Cr, Al