PECVD Recipes
Back to Vacuum Deposition Recipes.
PECVD 1 (PlasmaTherm 790)
SiN deposition (PECVD #1)
- SiN Standard Recipe
- [[media:|SiN Data (Deposition rate, Refractive index, Stress, HF etch rate)]]
- SiN Data March 2014
- SiN 1000A Thickness uniformity 2014
- Low Stress Si3N4 - Variable Stress Recipes
- Low Stress Si3N4 - Variable Stress Plot
SiO2 deposition (PECVD #1)
- SiO2 Standard Recipe
- SiO2 Data (Deposition rate, Refractive Index, Stress, HF etch rate)
- SiO2 Data March 2014
- SiO2 1000A Thickness uniformity 2014
SiOxNy deposition (PECVD #1)
PECVD 2 (Advanced Vacuum)
SiN deposition (PECVD #2)
- Nitride2 Standard Recipe
- Nitride2 Data (Deposition Rate, Refractive Index, Stress, HF etch rate)
- Nitride2 Data April 2014
- Nitride2 Thickness uniformity 2014
- Deposition Rate: ≈ 9.0 nm/min
- Refractive Index: ≈ 1.952
- Stress ≈ 484MPa
SiO2 deposition (PECVD #2)
- Oxide Standard Recipe
- Oxide Data (Deposition Rate, Refractive Index, Stress, HF etch rate )
- Oxide Data April 2014
- Oxide Thickness uniformity 2014
- Deposition Rate: ≈ 28.0 nm/min
- Refractive Index: ≈ 1.472
- Stress ≈ -270MPa
LS SiN deposition (PECVD #2)
- LS Nitride2 Standard Recipe
- LS Nitride2 Data (Deposition Rate, Refractive Index, Stress, HF etch rate )
- LS Nitride2 Data April 2014
- LS Nitride2 Thickness uniformity 2014
- Deposition Rate: ≈ 8.34 nm/min
- Refractive Index: ≈ 1.932
- Stress ≈ -45MPa
Amorphous-Si deposition (PECVD #2)
ICP-PECVD (Unaxis VLR)
SiN deposition (Unaxis VLR)
SiN (2% SiH4)
SiN (2% SiH4 - No-Ar)
50° | 100° | 250° |
---|---|---|
SiN (2% SiH4 - No-Ar - Extra N2)
50° | 100° | 250° |
---|---|---|
SiN (100% SiH4 )
50° | 100° | 250° |
---|---|---|
SiO2 deposition (Unaxis VLR)
SiO2 (2% SiH4)
50° | 100° | 250° |
---|---|---|
SiO2 (2% SiH4 - No Ar)
50° | 100° | 250° |
---|---|---|
SiO2 (100% SiH4 HDR)
50° | 100° | 250° |
---|---|---|
SiO2 (100% SiH4 LDR)
50° | 100° | 250° |
---|---|---|
Amorphous Si (100%SiH4 Ar He)
90° | 250° |
---|---|