Difference between revisions of "PECVD 1 (PlasmaTherm 790)"
(46 intermediate revisions by 5 users not shown) | |||
Line 2: | Line 2: | ||
|picture=PECVD1.jpg |
|picture=PECVD1.jpg |
||
|type = Vacuum Deposition |
|type = Vacuum Deposition |
||
− | |super= |
+ | |super= Don Freeborn |
− | |phone=(805)839- |
+ | |phone=(805)839-7975 |
|location=Bay 3 |
|location=Bay 3 |
||
− | |email= |
+ | |email=dfreeborn@ucsb.edu |
|description = PECVD Plasma Therm 790 For Oxides And Nitrides |
|description = PECVD Plasma Therm 790 For Oxides And Nitrides |
||
|manufacturer = Plasma-Therm |
|manufacturer = Plasma-Therm |
||
|materials = |
|materials = |
||
|toolid=16 |
|toolid=16 |
||
− | }} |
+ | }} |
+ | |||
+ | __TOC__ |
||
= About = |
= About = |
||
Line 29: | Line 31: | ||
=Documentation= |
=Documentation= |
||
+ | *[https://wiki.nanotech.ucsb.edu/w/images/d/d2/PECVD1_Operating_Instructions.pdf Operating Instuctions] |
||
− | *[[media:PECVD 1-Operating-Manual.pdf|Operating Instruction Manual]] |
||
+ | *[[PECVD1 Wafer Coating Process|Wafer Coating Process Traveler]] |
||
− | *[[media:PECVD 1 Standard recipes.pdf|PECVD Standard Recipes]] |
||
+ | *For particle counting method, see the [https://wiki.nanotech.ucsb.edu/wiki/Wafer_scanning_process_traveler Surfscan Scanning Procedure] |
||
− | *[[media:PECVD 1 SiO2 data-particles, thickness, index.pdf|PECVD1 SiO2 data-particles, thickness, index]] |
||
− | *[[media:PECVD 1 SiN data-particles, thickness, index.pdf|PECVD1 SiN data-particles, thickness, index]] |
Revision as of 16:21, 22 April 2020
|
About
This is a Plasma-Therm model 790 plasma enhanced chemical vapor deposition system for depositing SiO2, Si3N4, or SiOxNy dielectric films. The system uses a capacitively-coupled 13.56 MHz source excitation to produce the plasma between two parallel aluminum plates. The gas is injected over the sample through a 6” diameter showerhead. The samples are placed on the system anode (to minimize ion damage) which is heated to 250-350°C. SiO2 is produced from SiH4/He 2%/98% and N2O at 250°C. The typical deposition rate is 400 A/min. at 300 mT pressure. The typical BOE etch rate of this oxide is about 400 nm/min. Si3N4 is produced from SiH4/He 2%/98% and NH3 at 250°C or 350°C. The more dense films are produced at 350°C. The stress of the nitride can be altered by adjusting the N2:He ratio of the deposition. CF4/O2 plasmas are used to clean the chamber between depositions.
These films are typically used for capacitor dielectrics, chemical passivation layers, electrical insulators, reactive ion etching masks, and optical anti-reflective coatings. The system is fully programmable with windows-based software and has a wide array of pre-defined thicknesses. Custom programs for dielectric stacks or different process parameters can be written and saved.
Detailed Specifications
- Gases used: NH3, N2O, 2%SiH4/He, N2,CF4 and O2
- ~ 10mT ultimate chamber pressure
- 13.56 Mhz excitation freq.
- Sample size: pieces to 6” wafers
- Automatic tuning network
- RF Power control
- Full computer operation
- Standard recipes for a variety of film thicknesses
Documentation
- Operating Instuctions
- Wafer Coating Process Traveler
- For particle counting method, see the Surfscan Scanning Procedure