+ | [[Table SiO2 recipe]] |
||
− | '''PECVD#1 deposition - 300nm SiN film @250°C''' |
||
− | The wafers are ordered from SVM. These are low particle count 4" Si wafer where particle count is very low <100. |
||
− | a)Log in to PECVD #1 (Staff account) |
||
− | b)Edit standard recipes (for seasoning, deposition, and cleaning). You can '''ONLY''' change the time in recipes. |
||
− | • Seasoning recipe name: |
||
− | • Deposition recipe name: |
||
− | • Cleaning recipe name: |
||
− | c)Load the recipe for seasoning, and run it. The goal of this step is to coat chamber walls and prepare it for deposition. |
||
− | d)Vent the chamber and load 4"Si wafer (place in in the center of platen). You can place small pieces around, to protect wafer from moving |
||
− | e)Pump down and load the recipe for deposition. |
||
− | f) Edit the recipe for deposition, change the time ONLY. |
||
− | d) Run deposition |
||
− | e) Unload wafer |
||
− | f) Wipe sidewall first with wafter then with IPA. |
||
− | g) Load recipe for cleaning. Edit the recipe and etner required time for cleaning. |
||
− | h)Run the cleaning |