Difference between revisions of "Oxford ICP Etcher - Process Control Data"
(→Data - InP Ridge Etch (Oxford ICP Etcher): updated SEM 1) |
(updated) |
||
(40 intermediate revisions by 4 users not shown) | |||
Line 1: | Line 1: | ||
==Data - InP Ridge Etch (Oxford ICP Etcher)== |
==Data - InP Ridge Etch (Oxford ICP Etcher)== |
||
+ | [[PECVD1-(PlasmaTherm 790)|PECVD SiO2]] hardmask, patterned on [[Stepper 1 (GCA 6300)|Stepper #2 (AutoStep 200)]] & [[ICP Etch 1 (Panasonic E626I)|Panasonic ICP #1]] |
||
− | <br /> |
||
− | {{WIP}} |
||
{| class="wikitable" |
{| class="wikitable" |
||
− | | colspan="6" |'''InP Ridge Etch''': 3mT, 800W/65W, Cl2=18, H2=15, CH4=10sccm, time= |
+ | | colspan="6" |'''InP Ridge Etch''': 60°C, 3mT, 800W/65W, Cl2=18, H2=15, CH4=10sccm, time=5min00sec (300sec) |
+ | <u>Sample Size</u>: 2” unpatterned InP wafer “dummy” pieces, surrounding 1x1cm patterned sample, ~30-40% SiO<sub>2</sub> masking (NingC's pattern). Silicon carrier, no adhesive. |
||
+ | |||
+ | <u>Seasoning</u>: 10min “Std Clean - O2/SF6 Chamber Clean 20C (Edit Time)”, then load 2” “dummy” unpatterned InP wafer pieces on Si carrier (rough side up) and run 5min of “Std InP Ridge Etch - Cl2/CH4/H2 60C” |
||
+ | |||
+ | <u>Sample Prep</u>: Prior to dry etching, the sample is submerged in NH4OH:DI solution (XYZ:XYZ) for XYZ min, then DI rinsed and N2 dried. |
||
|- |
|- |
||
− | + | !'''Date''' |
|
− | + | !'''Sample#''' |
|
− | + | !'''Etch Rate (nm/min)''' |
|
− | + | !'''Etch Selectivity (InP/SiO2)''' |
|
− | + | !'''Comments''' |
|
− | + | !'''SEM Images''' |
|
+ | |- |
||
+ | |8/1/22 |
||
+ | |ND_60c_080122 |
||
+ | |298 |
||
+ | |9.9 |
||
+ | | |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/2/2e/CS_002.jpg <nowiki>[CS]</nowiki>] [https://wiki.nanotech.ucsb.edu/w/images/8/82/45D_001.jpg <nowiki>[45D]</nowiki>] |
||
+ | |- |
||
+ | |7/1/22 |
||
+ | |ND_60c_070122 |
||
+ | |294 |
||
+ | |11.9 |
||
+ | |Did not dip in NH4OH |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/6/65/45D_001_8-1-22.jpg <nowiki>[45D]</nowiki>] [https://wiki.nanotech.ucsb.edu/w/images/f/fb/A734DC5E-639E-4FDC-8F30-6F463470BC27.jpeg <nowiki>[CS]</nowiki>] |
||
+ | |- |
||
+ | |5/19/22 |
||
+ | |DJ_60c_007 |
||
+ | |326 |
||
+ | |TBA |
||
+ | |After chamber vented, cleaned (wet + dry). Forgot NH<sub>4</sub>OH wet-etch before dry etch. ER and profile look nominal. |
||
+ | |[[Media:Oxford InP 60C 45DEG02.jpg|[45°1]]],[[Media:Oxford InP 60C 45DEG05.jpg|[45°2]]] |
||
+ | [[Media:Oxford InP 60C XS05.jpg|[XS]]] |
||
+ | |- |
||
+ | |5/11/22 |
||
+ | |NP_60c_010 |
||
+ | |322 |
||
+ | |9.76 |
||
+ | | |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/2/25/Oxford_60c_10_45D_003.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/a/aa/Oxford_60c_10_CS_008.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | |5/4/22 |
||
+ | |NP_60c_009 |
||
+ | |318 |
||
+ | |10.5 |
||
+ | | |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/e/e8/InP_Oxford_60c_09_45D_005.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/3/3a/InP_Oxford_60c_09_CS_005.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | |4/27/22 |
||
+ | |NP_60c_008 |
||
+ | |320 |
||
+ | |9.2 |
||
+ | |directly after chamber clean. Identical results to before chamber clean. Chamber appeared clean when opened. |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/b/b4/Oxford_60c_08_45D_003.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/8/8d/Oxford_60c_08_CS_009.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | |4/26/22 |
||
+ | |NP_60c_007 |
||
+ | |312 |
||
+ | |9.93 |
||
+ | |day before chamber clean |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/6/66/Oxford_60c_07_45D_002.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/2/25/Oxford_60c_07_CS_001.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | |4/19/22 |
||
+ | |NP_60c_006 |
||
+ | |330 |
||
+ | |9.53 |
||
+ | |Chamber opened and brown gunk observed in chamber right after the etch was done. Etch rate unaffected. |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/0/03/Oxford_60c_06_45D_002.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/7/7d/Oxford_60c_06_CS_005.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | |4/13/22 |
||
+ | |NP_60c_005 |
||
+ | |320 |
||
+ | |11.51 |
||
+ | |New mask pattern with long lines to cleave through. |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/1/12/Oxford_60c_05_45D_004.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/5/56/Oxford_60c_05_CS_002.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | ! |
||
+ | |''↑ Changed mask pattern going forward (~50% open area), apparent etch rate changes due to measuring different features.'' |
||
+ | ! |
||
+ | |- |
||
+ | |3/30/22 |
||
+ | |NP_60c_004 |
||
+ | |427 |
||
+ | |11.17 |
||
+ | |*etched for 3min* |
||
+ | ~30-40% SiO<sub>2</sub> masking (NingC's pattern) |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/8/8a/Oxford_60c_04_45D_001.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/0/0a/Oxford_60c_04_CS_004.jpg <nowiki>[XS]</nowiki>] |
||
+ | |- |
||
+ | |1/26/22 |
||
+ | |NP_1_26_003 |
||
+ | |452 |
||
+ | |12.9 |
||
+ | |~30-40% SiO<sub>2</sub> masking (NingC's pattern) |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/e/ef/Oxford_Cal_01_26_22_003New_45d_001.jpg <nowiki>[45°]</nowiki>] |
||
+ | [https://wiki.nanotech.ucsb.edu/w/images/4/40/Oxford_Cal_01_26_22_031New_CS_001.jpg <nowiki>[XS]</nowiki>] |
||
+ | |} |
||
+ | |||
+ | ===Dependence on Sample Size=== |
||
+ | We have found that the size of the InP piece loaded (no adhesive, onto Silicon carrier wafer) affects the etch rate, but does not affect the etch profile - ie. etches are still smooth and vertical, but rate varies with sample area. |
||
+ | {| class="wikitable" |
||
+ | | colspan="8" |'''InP Ridge Etch''': 60°C, 3mT, 800W/65W, Cl2=18, H2=15, CH4=10sccm, time=5min05sec (305sec) |
||
+ | Silicon carrier, no adhesive. |
||
+ | |- |
||
+ | !Date |
||
+ | !Sample# |
||
+ | !Sample Size (dimensions, mm) |
||
+ | !Sample Size (area, mm<sup>2</sup>) |
||
+ | !Etch Rate (nm/min) |
||
+ | !Etch Selectivity (InP/SiO2) |
||
+ | !Comments |
||
+ | !SEM Images |
||
|- |
|- |
||
|1/11/22 |
|1/11/22 |
||
− | |<small> |
+ | |<small>DJ_InPRidge</small> |
+ | |4.5 x 2.5 |
||
− | |0.602 |
||
+ | |11.25 |
||
+ | |602 |
||
|64.6nm left |
|64.6nm left |
||
+ | |<small>~50% SiO<sub>2</sub> masking (GCA Calibration pattern)</small> |
||
− | |Software timing bugs fixed - new etch rate will appear slightly higher. |
||
|[https://wiki.nanotech.ucsb.edu/w/images/9/90/2022-01-11_Oxford_InP_Ridge_Wiki_Post_-_Cal01.pdf] |
|[https://wiki.nanotech.ucsb.edu/w/images/9/90/2022-01-11_Oxford_InP_Ridge_Wiki_Post_-_Cal01.pdf] |
||
|- |
|- |
||
|1/12/22 |
|1/12/22 |
||
− | |<small> |
+ | |<small>DJ_InPRidge</small> |
− | | |
+ | |4.5 x 3 |
+ | |13.5 |
||
+ | |563 |
||
|76.4nm left |
|76.4nm left |
||
+ | |<small>~50% SiO<sub>2</sub> masking (GCA Calibration pattern)</small> |
||
− | | |
||
− | |[https://wiki.nanotech.ucsb.edu/w/images/ |
+ | |[https://wiki.nanotech.ucsb.edu/w/images/d/d1/2022-01-12_Oxford_InP_Ridge_Wiki_Post_-_Cal02.pdf] |
|- |
|- |
||
|1/12/22 |
|1/12/22 |
||
− | |<small> |
+ | |<small>DJ_InP#3</small> |
− | | |
+ | |4.5 x 3 |
+ | |13.5 |
||
− | |80.1nm left |
||
+ | |612 |
||
+ | |71nm left |
||
+ | |<small>~50% SiO<sub>2</sub> masking (GCA Calibration pattern)</small> |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/f/ff/2022-01-12_Oxford_InP_Ridge_Wiki_Post_-_Cal03.pdf] |
||
+ | |- |
||
+ | |1/26/22 |
||
+ | |<small>NP_?</small> |
||
+ | |10 x 10 |
||
+ | |100 |
||
+ | |400-450 |
||
+ | |~250nm left |
||
+ | |<small>~30-40% SiO<sub>2</sub> masking (NingC's pattern)</small> |
||
| |
| |
||
+ | |- |
||
− | |[https://wiki.nanotech.ucsb.edu/w/images/2/26/I2210508.pdf] (wrong) |
||
+ | |1/26/22 |
||
− | |}<br /> |
||
+ | |<small>NP_?</small> |
||
+ | |1/4 of 50mm wafer |
||
+ | |490 |
||
+ | |378 |
||
+ | |276nm left |
||
+ | |<small>~30-40% SiO<sub>2</sub> masking (NingC's pattern)</small> |
||
+ | |[https://wiki.nanotech.ucsb.edu/w/images/c/c5/Oxford_Cal_01_26_22_New002_CS_002.jpg <nowiki>[1]</nowiki>] |
||
+ | |} |
Revision as of 17:25, 12 August 2022
Data - InP Ridge Etch (Oxford ICP Etcher)
PECVD SiO2 hardmask, patterned on Stepper #2 (AutoStep 200) & Panasonic ICP #1
InP Ridge Etch: 60°C, 3mT, 800W/65W, Cl2=18, H2=15, CH4=10sccm, time=5min00sec (300sec)
Sample Size: 2” unpatterned InP wafer “dummy” pieces, surrounding 1x1cm patterned sample, ~30-40% SiO2 masking (NingC's pattern). Silicon carrier, no adhesive. Seasoning: 10min “Std Clean - O2/SF6 Chamber Clean 20C (Edit Time)”, then load 2” “dummy” unpatterned InP wafer pieces on Si carrier (rough side up) and run 5min of “Std InP Ridge Etch - Cl2/CH4/H2 60C” Sample Prep: Prior to dry etching, the sample is submerged in NH4OH:DI solution (XYZ:XYZ) for XYZ min, then DI rinsed and N2 dried. | |||||
Date | Sample# | Etch Rate (nm/min) | Etch Selectivity (InP/SiO2) | Comments | SEM Images |
---|---|---|---|---|---|
8/1/22 | ND_60c_080122 | 298 | 9.9 | [CS] [45D] | |
7/1/22 | ND_60c_070122 | 294 | 11.9 | Did not dip in NH4OH | [45D] [CS] |
5/19/22 | DJ_60c_007 | 326 | TBA | After chamber vented, cleaned (wet + dry). Forgot NH4OH wet-etch before dry etch. ER and profile look nominal. | [45°1],[45°2] |
5/11/22 | NP_60c_010 | 322 | 9.76 | [45°] | |
5/4/22 | NP_60c_009 | 318 | 10.5 | [45°] | |
4/27/22 | NP_60c_008 | 320 | 9.2 | directly after chamber clean. Identical results to before chamber clean. Chamber appeared clean when opened. | [45°] |
4/26/22 | NP_60c_007 | 312 | 9.93 | day before chamber clean | [45°] |
4/19/22 | NP_60c_006 | 330 | 9.53 | Chamber opened and brown gunk observed in chamber right after the etch was done. Etch rate unaffected. | [45°] |
4/13/22 | NP_60c_005 | 320 | 11.51 | New mask pattern with long lines to cleave through. | [45°] |
↑ Changed mask pattern going forward (~50% open area), apparent etch rate changes due to measuring different features. | |||||
3/30/22 | NP_60c_004 | 427 | 11.17 | *etched for 3min*
~30-40% SiO2 masking (NingC's pattern) |
[45°] |
1/26/22 | NP_1_26_003 | 452 | 12.9 | ~30-40% SiO2 masking (NingC's pattern) | [45°] |
Dependence on Sample Size
We have found that the size of the InP piece loaded (no adhesive, onto Silicon carrier wafer) affects the etch rate, but does not affect the etch profile - ie. etches are still smooth and vertical, but rate varies with sample area.
InP Ridge Etch: 60°C, 3mT, 800W/65W, Cl2=18, H2=15, CH4=10sccm, time=5min05sec (305sec)
Silicon carrier, no adhesive. | |||||||
Date | Sample# | Sample Size (dimensions, mm) | Sample Size (area, mm2) | Etch Rate (nm/min) | Etch Selectivity (InP/SiO2) | Comments | SEM Images |
---|---|---|---|---|---|---|---|
1/11/22 | DJ_InPRidge | 4.5 x 2.5 | 11.25 | 602 | 64.6nm left | ~50% SiO2 masking (GCA Calibration pattern) | [1] |
1/12/22 | DJ_InPRidge | 4.5 x 3 | 13.5 | 563 | 76.4nm left | ~50% SiO2 masking (GCA Calibration pattern) | [2] |
1/12/22 | DJ_InP#3 | 4.5 x 3 | 13.5 | 612 | 71nm left | ~50% SiO2 masking (GCA Calibration pattern) | [3] |
1/26/22 | NP_? | 10 x 10 | 100 | 400-450 | ~250nm left | ~30-40% SiO2 masking (NingC's pattern) | |
1/26/22 | NP_? | 1/4 of 50mm wafer | 490 | 378 | 276nm left | ~30-40% SiO2 masking (NingC's pattern) | [1] |