Difference between revisions of "Logitech WBS7 - Procedure for Wax Mounting with bulk Crystalbond Stick"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
m (unload step)
(updated open/close procedure)
Line 2: Line 2:
 
Procedure for wax-mounting a 100mm Silicon wafer to another 100mm silicon carrier wafer.
 
Procedure for wax-mounting a 100mm Silicon wafer to another 100mm silicon carrier wafer.
 
* Place 4" polished handle wafer on hotplate set to 120°C
 
* Place 4" polished handle wafer on hotplate set to 120°C
* Apply approximately 1" diameter  size puddle of crystal bond wax from the stick. Then remove from hotplate. and cool. 
+
* Apply wax to center of wafer only, approximately <u>2-inch diameter</u> size puddle of crystal bond wax directly from the crystalbond stick.
  +
** Need a good amount of wax to ensure coverage to wafer edges, eg.<u>2-3mm thick</u>.
* Align work wafer on top of handle wafer placed on the glass plate of the Logitech bonder. Align the two wafers' flats.   Place 6" piece of filter paper on top of work wafer. 
 
  +
** Then remove from hotplate. and cool. 
* Turn on Logitech bonder (bottom right of machine chassis) 
 
  +
* Open Bonder chamber:
  +
** Unscrew clamp until higher than clamp arm,
  +
** Swing front of clamp arm to the left (Difficult to disengage),
  +
** Lift bonder lid (also takes some force to open - ignore vacuum reading on digital display)
  +
* On the glass plate of the Logitech bonder:
  +
** Align work wafer on top of handle wafer placed.
  +
** Align the two wafers' flats.  
  +
* Place 6" piece of filter paper on top of work wafer.
 
* Turn on Logitech bonder power (bottom right of machine chassis)
 
* Close and lock Logitech bonder hatch.
 
* Close and lock Logitech bonder hatch.
 
* On front panel of  Logitech bonder, confirm that Applied Pressure gauge is set to 5 psi.
 
* On front panel of  Logitech bonder, confirm that Applied Pressure gauge is set to 5 psi.
Line 28: Line 37:
 
** Unload when it displays ''Status: Ready''
 
** Unload when it displays ''Status: Ready''
 
* Turn off Logitech bonder (bottom right of machine chassis)
 
* Turn off Logitech bonder (bottom right of machine chassis)
  +
** Close lid and clamp

Revision as of 11:21, 11 August 2020

Contributed by Miguel Daal, Mazin Group  -- 2018-04-24

Procedure for wax-mounting a 100mm Silicon wafer to another 100mm silicon carrier wafer.

  • Place 4" polished handle wafer on hotplate set to 120°C
  • Apply wax to center of wafer only, approximately 2-inch diameter size puddle of crystal bond wax directly from the crystalbond stick.
    • Need a good amount of wax to ensure coverage to wafer edges, eg.2-3mm thick.
    • Then remove from hotplate. and cool. 
  • Open Bonder chamber:
    • Unscrew clamp until higher than clamp arm,
    • Swing front of clamp arm to the left (Difficult to disengage),
    • Lift bonder lid (also takes some force to open - ignore vacuum reading on digital display)
  • On the glass plate of the Logitech bonder:
    • Align work wafer on top of handle wafer placed.
    • Align the two wafers' flats.  
  • Place 6" piece of filter paper on top of work wafer.
  • Turn on Logitech bonder power (bottom right of machine chassis)
  • Close and lock Logitech bonder hatch.
  • On front panel of  Logitech bonder, confirm that Applied Pressure gauge is set to 5 psi.
  • Enter the process settings:
    • [setup] -> [Temp]
      • Bonding: 120°C      
      • Endpoint: 65°C
      • Offset: 20°C           
      • Units: C
    • [setup] -> [Vac Thres]
      • Proceed if Below: 2.4E+01 mBar
      • Abort if Below: 3.2E+01 mBar
    • [setup] -> [Process Control]
      • Process: Heat + Outgas
      • Soak: 5min
      • Bond: 15min
      • O.G. Limit: 20min
      • P.P. del: 1min
    • [Setup] -> [Heads]
      • Head 1 On
  • Start Process: [Process] -> [Start]
  • Process takes about 45min
    • Unload when it displays Status: Ready
  • Turn off Logitech bonder (bottom right of machine chassis)
    • Close lid and clamp