+ | ==='''<big>Photolithography Processes</big>'''=== |
+ | |||
#'''UV Optical Lithography''' |
#'''UV Optical Lithography''' |
||
#*[[#PositivePR |'''Stocked Lithography Chemical + Datasheets''']] |
#*[[#PositivePR |'''Stocked Lithography Chemical + Datasheets''']] |
||
#**''Lists all stocked photolith. chemicals, PRs, strippers, developers, and links to the chemical's application notes/datasheet, which detail the spin curves and nominal processes.'' |
#**''Lists all stocked photolith. chemicals, PRs, strippers, developers, and links to the chemical's application notes/datasheet, which detail the spin curves and nominal processes.'' |
||
#*[[#Photolithography_Recipes |'''Photo Lithography Recipe section''']] |
#*[[#Photolithography_Recipes |'''Photo Lithography Recipe section''']] |
||
− | #**''Starting recipes (spin, bake, exposure, develop etc.) for all photolith. tools.'' |
+ | #**''Starting recipes (spin, bake, exposure, develop etc.) for all photolith. tools.'' |
#**''Substrate/surface materials/pattern size can affect process parameters. Users may need to run Focus/Exposure Arrays/Matrix (FEA's/FEM's) with these processes to achieve high-resolution.'' |
#**''Substrate/surface materials/pattern size can affect process parameters. Users may need to run Focus/Exposure Arrays/Matrix (FEA's/FEM's) with these processes to achieve high-resolution.'' |
||
#**[[Contact Alignment Recipes|<u>Contact Aligner Recipes</u>]] |
#**[[Contact Alignment Recipes|<u>Contact Aligner Recipes</u>]] |
||
Line 23: | Line 24: | ||
#***[[Lithography Recipes#E-Beam Lithography Recipes|JEOL JBX-6300FS EBL]] |
#***[[Lithography Recipes#E-Beam Lithography Recipes|JEOL JBX-6300FS EBL]] |
||
#***[[Lithography Recipes#FIB Lithography Recipes .28Raith Velion.29|Raith Velion FIB]] |
#***[[Lithography Recipes#FIB Lithography Recipes .28Raith Velion.29|Raith Velion FIB]] |
||
+ | #[[Lithography Recipes#General Photolithography Techniques|'''General Photolithography Techniques''']] |
||
+ | #*''Techniques for improving litho. or solving common photolith. problems.'' |
||
#'''[[Lithography Recipes#Lift-Off Recipes|Lift-Off Recipes]]''' |
#'''[[Lithography Recipes#Lift-Off Recipes|Lift-Off Recipes]]''' |
||
#*''Verified Recipes for lift-off using various photolith. tools'' |
#*''Verified Recipes for lift-off using various photolith. tools'' |
||
Line 32: | Line 35: | ||
#**''Provided for reference, also showing starting recipes and usage info.'' |
#**''Provided for reference, also showing starting recipes and usage info.'' |
||
#'''[[Lithography Recipes#Holography Recipes|Holography]]''' |
#'''[[Lithography Recipes#Holography Recipes|Holography]]''' |
||
− | #*''For 1-D and 2-D gratings with 220nm nominal period, available on substrates up to 1 inch square.'' |
+ | #*''For 1-D and 2-D gratings with 220nm nominal period, available on substrates up to 1 inch square.'' |
#*''Recipes for silicon substrates are provided, and have been translated to other substrates by users.'' |
#*''Recipes for silicon substrates are provided, and have been translated to other substrates by users.'' |
||
− | #*''Datasheets are provided with starting recipes and usage info.'' |
+ | #*''Datasheets are provided with starting recipes and usage info.'' |
|- |
|- |
||
| |
| |
||
− | === |
+ | ==='''<big>Photolithography Chemicals/Materials</big>'''=== |
+ | |||
#'''[[#Underlayers |Underlayers]]''' |
#'''[[#Underlayers |Underlayers]]''' |
||
− | #*''These are used beneath resists for both adhesive purposes and to enable bi-layer lift-off profiles for use with photoresist.'' |
+ | #*''These are used beneath resists for both adhesive purposes and to enable bi-layer lift-off profiles for use with photoresist.'' |
#*''Datasheets are provided.'' |
#*''Datasheets are provided.'' |
||
#'''[[#AntiReflectionCoatings |Anti-Reflection Coatings]]''': |
#'''[[#AntiReflectionCoatings |Anti-Reflection Coatings]]''': |
||
#*[[#Photolithography_Recipes |The Photoresist Recipes]] section contains recipes using these materials. |
#*[[#Photolithography_Recipes |The Photoresist Recipes]] section contains recipes using these materials. |
||
− | #*''Bottom Anti-Reflection Coatings (BARC) are used in the stepper systems, underneath the resists to eliminate substrate reflections that can affect resolution and repeatability for small, near resolution limited, feature sizes.'' |
+ | #*''Bottom Anti-Reflection Coatings (BARC) are used in the stepper systems, underneath the resists to eliminate substrate reflections that can affect resolution and repeatability for small, near resolution limited, feature sizes.'' |
#*''Datasheets are provided for reference on use of the materials.'' |
#*''Datasheets are provided for reference on use of the materials.'' |
||
#'''[[#ContrastEnhancement |Contrast Enhancement Materials (CEM)]]''' |
#'''[[#ContrastEnhancement |Contrast Enhancement Materials (CEM)]]''' |
||
Line 51: | Line 55: | ||
#*''Datasheets provided with usage info.'' |
#*''Datasheets provided with usage info.'' |
||
#'''[[#AdhesionPromoters |Adhesion Promoters]]''' |
#'''[[#AdhesionPromoters |Adhesion Promoters]]''' |
||
− | #*''These are used to improve wetting of photoresists to your substrate.'' |
+ | #*''These are used to improve wetting of photoresists to your substrate.'' |
#*''Datasheets are provided on use of these materials.'' |
#*''Datasheets are provided on use of these materials.'' |
||
#'''[[#SpinOnDielectrics |Low-K Spin-on Dielectrics]]''' |
#'''[[#SpinOnDielectrics |Low-K Spin-on Dielectrics]]''' |
||
Line 62: | Line 66: | ||
#*''Remover and Photoresist Strippers are used to dissolve PR during lift-off or after etching.'' |
#*''Remover and Photoresist Strippers are used to dissolve PR during lift-off or after etching.'' |
||
|} |
|} |
||
+ | |||
+ | ==General Photolithography Techniques== |
||
+ | |||
+ | *[[Photolithography - Improving Adhesion Photoresist Adhesion|HMDS Process for Improving Adhesion]] |
||
+ | **''Use these procedures if you are finding poor adhesion PR lifting-off), or for chemicals (like BHF) that attack the PR adhesion interface strongly.'' |
||
+ | *[[Photolithography - Manual Edge-Bead Removal Techniques|Manual Edge-Bead Removal Techniques]] |
||
+ | **''Removing the edge-bead from your substrate will help with contact litho resolution and alignment.'' |
||
==Photolithography Recipes== |
==Photolithography Recipes== |
||
+ | * <small>'''R''': ''Recipe is available. Clicking this link will take you to the recipe.''</small> |
||
− | {{Recipe Table Explanation}} |
||
+ | * <small>'''A''': ''Material is available for use, but no recipes are provided.''</small> |
||
+ | |||
''Click the tool title to go to recipes for that tool.'' |
''Click the tool title to go to recipes for that tool.'' |
||
Line 74: | Line 87: | ||
|- |
|- |
||
− | | bgcolor="#EAECF0" | |
+ | | bgcolor="#EAECF0" |<!-- INTENTIONALLY BLANK --> |
! colspan="2" align="center" |'''[[Contact Alignment Recipes|<big>Contact Aligner Recipes</big>]]''' |
! colspan="2" align="center" |'''[[Contact Alignment Recipes|<big>Contact Aligner Recipes</big>]]''' |
||
! colspan="3" align="center" |'''[[Stepper Recipes|<big>Stepper Recipes</big>]]''' |
! colspan="3" align="center" |'''[[Stepper Recipes|<big>Stepper Recipes</big>]]''' |
||
Line 115: | Line 128: | ||
|- bgcolor="EEFFFF" |
|- bgcolor="EEFFFF" |
||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/8/8b/OCG825-Positive-Resist-Datasheet.pdf OCG 825-35CS] |
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/8/8b/OCG825-Positive-Resist-Datasheet.pdf OCG 825-35CS] |
||
− | | |
+ | |A |
− | | |
+ | |A |
− | | |
+ | |A |
− | | |
+ | |A |
| |
| |
||
|A |
|A |
||
Line 238: | Line 251: | ||
|- |
|- |
||
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/7/78/SU-8-2015-revA.pdf SU-8 2005,2010,2015] |
| bgcolor="#D0E7FF" align="center" |[//wiki.nanotech.ucsb.edu/w/images/7/78/SU-8-2015-revA.pdf SU-8 2005,2010,2015] |
||
− | | |
+ | |A |
− | | |
+ | |{{rl|Contact_Alignment_Recipes|Negative Resist (MA-6)}} |
− | | |
+ | |A |
− | | |
+ | |A |
| |
| |
||
|A |
|A |
||
Line 261: | Line 274: | ||
|A |
|A |
||
|- bgcolor="EEFFFF" |
|- bgcolor="EEFFFF" |
||
− | ! bgcolor="#D0E7FF" align="center" | |
+ | ! bgcolor="#D0E7FF" align="center" |'''Anti-Reflection Coatings''' |
{{LithRecipe Table}} |
{{LithRecipe Table}} |
||
|- |
|- |
||
Line 297: | Line 310: | ||
*{{fl|Liftoff-Techniques.pdf|Lift-Off Description/Tutorial}} |
*{{fl|Liftoff-Techniques.pdf|Lift-Off Description/Tutorial}} |
||
**How it works, process limits and considerations for designing your process |
**How it works, process limits and considerations for designing your process |
||
+ | *[[Lift-Off with I-Line Imaging Resist + LOL2000 Underlayer|I-Line Lift-Off: Bi-Layer Process with LOL2000 Underlayer]] |
||
+ | **''Single Expose/Develop process for simplicity'' |
||
+ | **''Up to ~130nm metal thickness & ≥500nm-1000nm gap between metal.'' |
||
+ | **''Can use any I-Line litho tool (GCA Stepper, Contact aligner, MLA)'' |
||
*{{fl|Bi-LayerContactprocesswithPMGI.pdf|I-Line Lift-Off: Bi-Layer Process with PMGI Underlayer and Contact Aligner}} |
*{{fl|Bi-LayerContactprocesswithPMGI.pdf|I-Line Lift-Off: Bi-Layer Process with PMGI Underlayer and Contact Aligner}} |
||
+ | **''Multiple processes for Metal thicknesses ~800nm to ~2.5µm'' |
||
+ | **''Uses multiple DUV Flood exposure/develop cycles to create undercut.'' |
||
+ | **''Can be transferred to other I-Line litho tools (Stepper, MLA etc.)'' |
||
*[[Lift-Off with DUV Imaging + PMGI Underlayer|DUV Lift-Off: UV6 Imaging Resist + PMGI Underlayer]] |
*[[Lift-Off with DUV Imaging + PMGI Underlayer|DUV Lift-Off: UV6 Imaging Resist + PMGI Underlayer]] |
||
+ | **''Single-expose/develop process'' |
||
+ | **''Up to ~65nm metal thickness & ~350nm gap between metal'' |
||
+ | **''Use thicker PMGI for thicker metals'' |
||
==[[E-Beam Lithography System (JEOL JBX-6300FS)|E-Beam Lithography Recipes (JEOL JBX-6300FS)]]== |
==[[E-Beam Lithography System (JEOL JBX-6300FS)|E-Beam Lithography Recipes (JEOL JBX-6300FS)]]== |
||
Line 304: | Line 327: | ||
*Under Development. |
*Under Development. |
||
− | == |
+ | ==[[Focused Ion-Beam Lithography (Raith Velion)|FIB Lithography Recipes (Raith Velion)]]== |
''To Be Added'' |
''To Be Added'' |
||
Line 314: | Line 337: | ||
==[[Holographic Lith/PL Setup (Custom)|Holography Recipes]]== |
==[[Holographic Lith/PL Setup (Custom)|Holography Recipes]]== |
||
''The Holography recipes here use the BARC layer XHRiC-11 & the high-res. I-Line photoresist THMR-IP3600HP-D.'' |
''The Holography recipes here use the BARC layer XHRiC-11 & the high-res. I-Line photoresist THMR-IP3600HP-D.'' |
||
+ | |||
*{{fl|Holography_Process_for_1D-lines_and_2D-dots_%28ARC-11_%26_THMR-IP3600HP-D%29-updated-4-8-2021.pdf|Standard Holography Process - on SiO2 on Si}} |
*{{fl|Holography_Process_for_1D-lines_and_2D-dots_%28ARC-11_%26_THMR-IP3600HP-D%29-updated-4-8-2021.pdf|Standard Holography Process - on SiO2 on Si}} |
||
*{{fl|Holography-Process-Variation-revA.pdf|Holography Process Variations - Set-up Angle - Etching into SiO2 and Si}} |
*{{fl|Holography-Process-Variation-revA.pdf|Holography Process Variations - Set-up Angle - Etching into SiO2 and Si}} |
Revision as of 15:48, 7 July 2022
Table of Contents |
---|
Photolithography Processes
|
Photolithography Chemicals/Materials
|
General Photolithography Techniques
- HMDS Process for Improving Adhesion
- Use these procedures if you are finding poor adhesion PR lifting-off), or for chemicals (like BHF) that attack the PR adhesion interface strongly.
- Manual Edge-Bead Removal Techniques
- Removing the edge-bead from your substrate will help with contact litho resolution and alignment.
Photolithography Recipes
- R: Recipe is available. Clicking this link will take you to the recipe.
- A: Material is available for use, but no recipes are provided.
Click the tool title to go to recipes for that tool.
Click the photoresist title to get the datasheet, also found in Stocked Chemicals + Datasheets.
Lift-Off Recipes
- Lift-Off Description/Tutorial
- How it works, process limits and considerations for designing your process
- I-Line Lift-Off: Bi-Layer Process with LOL2000 Underlayer
- Single Expose/Develop process for simplicity
- Up to ~130nm metal thickness & ≥500nm-1000nm gap between metal.
- Can use any I-Line litho tool (GCA Stepper, Contact aligner, MLA)
- I-Line Lift-Off: Bi-Layer Process with PMGI Underlayer and Contact Aligner
- Multiple processes for Metal thicknesses ~800nm to ~2.5µm
- Uses multiple DUV Flood exposure/develop cycles to create undercut.
- Can be transferred to other I-Line litho tools (Stepper, MLA etc.)
- DUV Lift-Off: UV6 Imaging Resist + PMGI Underlayer
- Single-expose/develop process
- Up to ~65nm metal thickness & ~350nm gap between metal
- Use thicker PMGI for thicker metals
E-Beam Lithography Recipes (JEOL JBX-6300FS)
- Under Development.
FIB Lithography Recipes (Raith Velion)
To Be Added
Automated Coat/Develop System Recipes (S-Cubed Flexi)
Recipes pre-loaded on the S-Cubed Flexi automated coat/bake/develop system. Only staff may write new recipes, contact the tool supervisor for more info.
To Be Added
Holography Recipes
The Holography recipes here use the BARC layer XHRiC-11 & the high-res. I-Line photoresist THMR-IP3600HP-D.
- Standard Holography Process - on SiO2 on Si
- Holography Process Variations - Set-up Angle - Etching into SiO2 and Si
- Etch SiO2 Nano-structure - Changing Side-wall Angle - Etching into Si with a different line-width
- Reduce SiO2 Nanowire Diameter - Thermal Oxidation - Vapor HF Etching
Low-K Spin-On Dielectric Recipes
Chemicals Stocked + Datasheets
The following is a list of the lithography chemicals we stock in the lab, with links to the datasheets for each. The datasheets will often have important processing info such as spin-speed vs. thickness curves, typical process parameters, bake temps/times etc.