Difference between revisions of "ICP Etch 1 (Panasonic E626I)"

From UCSB Nanofab Wiki
Jump to navigation Jump to search
(Updated Gas Change SOP)
 
(6 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{tool|{{PAGENAME}}
+
{{tool2|{{PAGENAME}}
 
|picture=ICP2.jpg
 
|picture=ICP2.jpg
 
|type = Dry Etch
 
|type = Dry Etch
 
|super= Lee Sawyer
 
|super= Lee Sawyer
 +
|super2= Don Freeborn
 
|location=Bay 2
 
|location=Bay 2
|description = ?
+
|description = ICP Etching and Ashing Multi-Chamber Tool
 
|manufacturer = Panasonic Factory Solutions, Japan
 
|manufacturer = Panasonic Factory Solutions, Japan
 
|materials =  
 
|materials =  
Line 27: Line 28:
 
Chamber three "Rinse Chamber" is a DI rinsing chamber that is not used/offline.  
 
Chamber three "Rinse Chamber" is a DI rinsing chamber that is not used/offline.  
  
The system accepts 6” wafers with SEMI-std. flats. Users often mount smaller pieces to the wafers, usually with easily removable oil to improve uniform heatsinking.  
+
The system accepts 6” wafers with SEMI-std. flats. Users often mount smaller pieces to the wafers, usually with easily removable oil to improve uniform heat-sinking.  
  
 
In Automatic mode, multiple wafers can be run through automatically with the cassette-based system.
 
In Automatic mode, multiple wafers can be run through automatically with the cassette-based system.
Line 41: Line 42:
 
*Optimal Emission Monitoring
 
*Optimal Emission Monitoring
 
*Etch pressure from 0.1 Pa to 5 Pa (0.75 mT - 37.5 mT)
 
*Etch pressure from 0.1 Pa to 5 Pa (0.75 mT - 37.5 mT)
*Cl<sub>2</sub>, BCl<sub>3</sub>, N<sub>2</sub>, O<sub>2,</sub> (CHF<sub>3</sub> or Ar), (CF<sub>4</sub> or SF<sub>6</sub>) in etch chamber
+
*Cl<sub>2</sub>, BCl<sub>3</sub>, O<sub>2,</sub> (CHF<sub>3</sub> or Ar), (CF<sub>4</sub> or SF<sub>6</sub>), and (N<sub>2</sub> or He) in etch chamber
 
*Room Temp - 80°C sample temperature for etching. Default 12°C Chuck Temperature.
 
*Room Temp - 80°C sample temperature for etching. Default 12°C Chuck Temperature.
 
*Laser Etch Monitoring: [[Laser Etch Monitoring|Intellemetrics LEP 500]]
 
*Laser Etch Monitoring: [[Laser Etch Monitoring|Intellemetrics LEP 500]]
Line 57: Line 58:
 
*[https://wiki.nanotech.ucsb.edu/w/images/0/0a/ICP_-1_Gas_Change_CF4-SF6-CF4.pdf]{{file|Panasonic 1 instructions.pdf|Panasonic _1_instructions.pdf}}
 
*[https://wiki.nanotech.ucsb.edu/w/images/0/0a/ICP_-1_Gas_Change_CF4-SF6-CF4.pdf]{{file|Panasonic 1 instructions.pdf|Panasonic _1_instructions.pdf}}
 
*[https://wiki.nanotech.ucsb.edu/w/images/0/0a/ICP_-1_Gas_Change_CF4-SF6-CF4.pdf ICP #1 Gas Change Procedure: CF<sub>4</sub>-SF<sub>6</sub>-CF<sub>4</sub>]
 
*[https://wiki.nanotech.ucsb.edu/w/images/0/0a/ICP_-1_Gas_Change_CF4-SF6-CF4.pdf ICP #1 Gas Change Procedure: CF<sub>4</sub>-SF<sub>6</sub>-CF<sub>4</sub>]
*ICP #1 Gas Change Procedure: CHF<sub>3</sub>-Ar-CHF<sub>3</sub>
+
*[https://wiki.nanotech.ucsb.edu/w/images/7/7a/ICP_-1_Gas_Change_CHF3-Ar-CHF3.pdf ICP #1 Gas Change Procedure: CHF<sub>3</sub>-Ar-CHF<sub>3</sub>]
*ICP #1 Gas Change Procedure: N<sub>2</sub>-He-N<sub>2</sub>
+
*[https://wiki.nanotech.ucsb.edu/w/images/f/fa/ICP_-1_Gas_Change_N2-He-N2.pdf ICP #1 Gas Change Procedure: N<sub>2</sub>-He-N<sub>2</sub>]
*ICP #1 Rules and Important Notes
+
*[https://wiki.nanotech.ucsb.edu/w/images/3/3a/ICP_-1_Rules_%26_Important_Notes.pdf ICP #1 Rules and Important Notes]
 +
*[https://wiki.nanotech.ucsb.edu/w/images/d/de/ICP_-1_Wafer_Type_Change.pdf ICP #1 Wafer Type Change Procedure]
 
*[[Laser Etch Monitoring|Laser Etch Monitor procedures]]
 
*[[Laser Etch Monitoring|Laser Etch Monitor procedures]]
  
Line 66: Line 68:
 
*[[ICP Etching Recipes#ICP Etch 1 .28Panasonic E626I.29|'''Recipes > Dry Etching > ICP Etch 1''']] page lists all qualified and contributed recipes for this tool.
 
*[[ICP Etching Recipes#ICP Etch 1 .28Panasonic E626I.29|'''Recipes > Dry Etching > ICP Etch 1''']] page lists all qualified and contributed recipes for this tool.
 
**Starting point recipes for ICP#1
 
**Starting point recipes for ICP#1
 +
*[[ICP Etching Recipes#Process Control Data .28Panasonic 1.29|Process Control Data]]
 
**''Historical Data'' records "calibration" etches to test tool performance.
 
**''Historical Data'' records "calibration" etches to test tool performance.
 
*The [[Dry Etching Recipes|'''Recipes > Dry Etching Recipes''']]  
 
*The [[Dry Etching Recipes|'''Recipes > Dry Etching Recipes''']]  
**Master table lists all contributed dry etches and materials across tools.
+
**Master table lists all contributed '''dry etches vs etched materials''' across tools.

Latest revision as of 09:45, 30 August 2022

ICP Etch 1 (Panasonic E626I)
ICP2.jpg
Location Bay 2
Tool Type Dry Etch
Manufacturer Panasonic Factory Solutions, Japan
Description ICP Etching and Ashing Multi-Chamber Tool

Primary Supervisor Lee Sawyer
(805) 893-2123
lee_sawyer@ucsb.edu

Secondary Supervisor

Don Freeborn


Recipes Dry Etch RecipesN/A

SignupMonkey: Sign up for this tool


About

This is a three-chamber tool for etching of a variety of materials.

Chamber one "Etch Chamber" is configured as an ICP etching tool with 1000 W ICP power, 500 W RF substrate power, and RT - 80°C operation with back-side He cooling and an electrostatic chuck to maintain controlled surface temperatures during etching.

This chamber has the following dedicated gas sources: Cl2, BCl3, and O2

The chamber also has the following gas sources, where two of the lines must be manually switched between the two options shown (gasses can't be used simultaneously): N2/He, CHF3/Ar and CF4 /SF6

The system can be used to etch a variety of materials from SiO2 to metals to compound semiconductors. The chamber is evacuated with a 2000 lpm Osaka Vacuum magnetically levitated turbo pump, and is load-locked for fast pump down.

The in-situ laser monitor installed on the etch chamber allows for repeatable etches and endpoint detection via continuous optical monitoring of the wafer reflectivity in a user-determined location, through a porthole on the chamber.

Chamber two, "Ashing Chamber" is a 2000 W ICP chamber configures for plasma "ashing" of photoresist and other materials such as BCB. The substrate is not biased for isotropic etching, and the chamber has CF4 and O2 for the gases. This is especially well-suited for omni-directional etching of photoresist/PR removal, or BCB etch-back.

Chamber three "Rinse Chamber" is a DI rinsing chamber that is not used/offline.

The system accepts 6” wafers with SEMI-std. flats. Users often mount smaller pieces to the wafers, usually with easily removable oil to improve uniform heat-sinking.

In Automatic mode, multiple wafers can be run through automatically with the cassette-based system.

Detailed Specifications

  • 1000 W ICP source, 500 W RF Sample Bias Source in etching chamber
  • Multiple 6” diameter wafer capable system
  • Pieces possible by mounting to 6” wafer

Etch Chamber:

  • Optimal Emission Monitoring
  • Etch pressure from 0.1 Pa to 5 Pa (0.75 mT - 37.5 mT)
  • Cl2, BCl3, O2, (CHF3 or Ar), (CF4 or SF6), and (N2 or He) in etch chamber
  • Room Temp - 80°C sample temperature for etching. Default 12°C Chuck Temperature.
  • Laser Etch Monitoring: Intellemetrics LEP 500

Ashing Chamber:

  • 2000 W ICP ashing chamber
  • RT - 250°C sample temperature for ashing
  • Ashing pressures 50 mT - 500 mT
  • O2, N2, CF4, H2O Vapor for ashing chamber
  • Room Temp. to 270°C etching. Default 50°C.

Documentation