Fluorescence Microscope (Olympus MX51)

From UCSB Nanofab Wiki
Revision as of 11:07, 15 April 2020 by Silva (talk | contribs)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Fluorescence Microscope (Olympus MX51)
Fluorescence.jpg
Tool Type Inspection, Test and Characterization
Location Bay 6
Description Optical Microscope
Manufacturer Olympus
Model MX51


About

System for general purpose imaging. The system has bright field, dark-field, DIC imaging mode and Three fluorescence filters. Screen captures and calibrated measurements can be made through use of native Olympus' Streams Essential Software, installed on the attached computer.

Detailed Specifications

  • Native Olympus Streams Essentials software & Camera:
    • Digital image capture & calibrated measurement (calibrations locked)
  • Objectives: 5x, 10x, 20x, 50x, 100x, 150x
  • Reflected (Episcopic) & Transmitted (Diascopic) Illumination modes
  • Filters via rotatable selector:
    • BF: Bright Field normal top-illuminated imaging
    • DF: Dark-Field imaging
    • DIC: Differential Interference Contrast "Nomarski" Imaging. Also requires "DIC-R" prism to be inserted. Polarization manually selected on DIC-R prism.
  • Fluorescence Filters (see Olympus tutorial):
    • You must contact Tony Bosch for training on Fluorescence before using this feature!!!
    • WB: "Blue" Fluorescence Filter. Excitation: 450-480nm / Dichroic 500nm / Barrier 515nm+
    • WG: "Green" Fluorescence Filter. Excitation: 510-550nm / Dichroic 570nm / Barrier 590nm+
    • WU: "Ultraviolet" Fluorescence Filter. Excitation: 330-385nm / Dichroic 400nm / Barrier 420nm+

Documentation