DUV Flood Expose

From UCSB Nanofab Wiki
Revision as of 09:19, 12 March 2014 by Silva m (talk | contribs)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
DUV Flood Expose
DUV.jpg
Tool Type Lithography
Location Bay 6
Supervisor Mike Silva
Supervisor Phone (805) 893-3096
Supervisor E-Mail silva@ece.ucsb.edu
Description Deep UV Flood Exposer / Light source and HA-1KC2 power supply
Manufacturer AB Manufacturing



About

This unit consists of a collimated deep ultraviolet (DUV) light source (mercury gas discharge short arc lamp) and power supply. The substrate is placed on a rotating chuck (not presently working) and is exposed by opening a timer-controlled shutter.

Materials that are exposed are primarily spun-on thin films such as PMMA, PMGI, etc.

Detailed Specifications

  • DUV wavelengths are 200-260 nm; the lamp power is limited to either 1250 or 2100 watts and can operate in either constant intensity or constant power mode
  • The full spectrum of Hg emission wavelengths from 200-450 nm is present on the sample
  • Exposure can be performed on a 4" wafer
  • Lamps are nominally rated for 400 hours
  • A reset timer was added in Feb. 1995 to limit "on" time (ie. provide for auto shut-off) for bulb life conservation

Spectra Images

Aligner with No Filter

Aligner with No Filter

Aligner with Filter for i-line Exposure

Aligner with Filter for i-line Exposure

DUV Exposure System Spectra

DUV Exposure System Spectra